64排螺旋CT轴扫节段扫描对降低腹部CT检查辐射剂量的作用 被引量:13

Radiation dose reduction with cine segment technique in abdominal CT scan
下载PDF
导出
摘要 目的研究轴扫节段扫描对降低腹部CT检查辐射剂量的作用。方法 54例患者按照就诊顺序分成3组,A组20例螺旋扫描,B组16例轴扫扫描,C组18例轴扫节段扫描,均使用120 kV,自动毫安,噪声指数=8,分别进行平扫和增强扫描,比较各组图像质量、噪声水平和平均辐射剂量。结果各组图像质量得分之间差异无统计学意义(P>0.05)。平扫3组图像噪声差异无统计学意义,增强后A组和B组图像噪声之间差异无统计学意义(P>0.05),C组和其他两组之间差异有统计学意义(P<0.01)。有效剂量C组比A组降低约47%,比B组降低约25%的,差异有统计学意义(P<0.01)。结论自动毫安技术和轴扫节段扫描能有效降低腹部CT扫描的辐射剂量,保证图像质量。 Objective To evaluate the value of low-dose multi-slice CT(MSCT) scanning with smart-mA and Cine Segment technique in radiation dose reduction in abdominal CT scan.Methods A total of 54 patients with abdominal CT scan were divided into three groups,Group A(Smart-mA scanning with 120 kV,n=20),Group B(Smart-mA with Noise Index(NI) =8;Cine scanning with 120 kV;n=16) and Group C(Smart-mA,NI=8;Cine segment scanning with 120 kV;n=8).Radiation dose,and quality score and noise of image were compared.Results There was no significant difference in image quality score among the 3 groups.Image noise in Group C was significant higher than those in the other 2 groups after contrast injection(P0.01).However,significant reduction of radiation dose by 47% revealed in Group C when comparing with that in Group A;and by 25% when comparing with that in Group A(P0.01).Conclusion Smart-mA and Cine Segment scanning techniques effectively reduce the radiation dose in abdominal CT scan.
出处 《广东医学》 CAS CSCD 北大核心 2012年第7期929-932,共4页 Guangdong Medical Journal
基金 广西自然科学基金资助项目(编号:2011GXNSFB018084)
关键词 体层摄影术 X线计算机 腹部扫描 低剂量 tomography X-ray computed abdominal CT low-dose
  • 相关文献

参考文献15

  • 1章士正.规范腹部CT检查保证诊断质量[J].中华放射学杂志,2007,41(9):898-899. 被引量:10
  • 2KALRA M K, PRADAS S, SAINI S, et al. Clinical comparison of standard- dose and 50% reduced- dose abdominal CT: effect on image quality[J]. AJR Am J Roentgenol, 2002, 179(5) : 1601 - 1606.
  • 3NAKAYAMA Y, AWAI K, FUNAMA Y, et al. Abdominal CT with low tube voltage: preliminary observations about radiatiol dose, contrast enhancement, image quality, and noise [ J ]. Radiology, 2005, 237(3): 945-951.
  • 4肖圣祥,毛定立,柴春华,阮凌翔,肖文波,周先勇.64排螺旋CT腹部血管成像低剂量扫描的可行性研究[J].中华放射医学与防护杂志,2010,30(4):480-482. 被引量:14
  • 5殷瑞根,吴树春,王冬青,彭卫斌,孙维斌,范志刚,廖谦,赵亮.低剂量多层螺旋CT在急腹症中的初步应用[J].临床放射学杂志,2005,24(7):609-612. 被引量:29
  • 6孙东辉,万业达,许相丰,张继扬,何岸苇,张琳,何珍.低剂量多层螺旋CT测量腹部脂肪的实验研究[J].临床放射学杂志,2010,29(8):1123-1125. 被引量:8
  • 7KALRA M K, RIZZO S M, NOVELLINE R A. Reducing radiation dose in emergency computed tomography with automatic exposure control techniques [ J ]. Emerg Radiol, 2005, 11 (5) : 267 - 274.
  • 8GREESS H, WOLF H, BAUM U, el at. Dose reduction in computed tomography by attenuation - based on - line modulation of tube current : evaluation of six anatomical regions[ J]. Eur Radiol, 2000, 10(2): 391 -394.
  • 9冯琦,龚红霞,庄治国,殷焱,沈云,高娜,许建荣.自动毫安低剂量CT扫描对结肠病变的显示及其对图像质量的影响[J].中华放射学杂志,2008,42(10):1022-1025. 被引量:35
  • 10FUNAMA Y, AWAI K, NAKAYAMA Y, et al. Radiation dose reduction without degradation of low - contrast detectability at abdominal muhisection CT with a low - tube voltage technique:phantom study [ J ]. Radiology, 2005, 237 (3) : 905 - 910.

二级参考文献69

共引文献200

同被引文献125

引证文献13

二级引证文献86

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部