利用Barracuda系统测量医用CT剂量指数量值不确定度评定 被引量:2

Evaluation of uncertainty in measurement of computed tomography dose index by Barracuda system
原文传递
导出
摘要 目的为保证医用CT剂量指数量值的准确性,评定其不确定度。方法建立Barracuda系统10 cm长杆电离室测量CT剂量指数数学模型,分析不确定度来源,通过计算给出不确定度。结果相对扩展不确定度为6.5%。结论利用Bar-racuda系统测量医用CT剂量指数量值不确定度主要受测量系统不确定度、校准因子、层厚偏差等分量的影响。 [Objective]In order to ensure the accuracy of computed tomography dose index,to evaluate the measurement uncertainty.[Methods]The mathematical model of Barracuda system with 10 cm long ionization chamber in measurement of computed tomography dose index was established,the uncertainty was calculated by analyzing the uncertainty source.[Results]The relative expanded uncertainty was 6.5%.[Conclusion]The uncertainty of Barracuda system in measurement of computed tomography dose index is mainly influenced by the uncertainty of the measurement system,the calibration factors and the thickness deviation.
出处 《职业与健康》 CAS 2012年第8期954-955,共2页 Occupation and Health
关键词 Barracuda、医用CT 剂量指数 测量不确定度 Barracuda Computed tomography Dose index Measurement uncertainty