期刊文献+

多普勒超声检测动脉栓子的实验研究 被引量:1

An experimental study on detection of arterial emboli using Doppler ultrasound
下载PDF
导出
摘要 目的 为了观察多普勒超声检测动脉气泡及固体栓子的可能性。方法 采用 2MHz脉冲波多普勒探头在兔腹主动脉检测由左肾动脉输入的气泡栓子和兔全血凝块栓子。结果 不管是气泡栓子还是全血凝块栓子在多普勒频谱图上均产生相似的高强度短暂信号。结论 多普勒超声能够检测动脉血流中气泡和固体栓子 ,但不能区分是气泡栓子还是固体栓子。 Objective This study was designed to test the ability of Doppler ultrasound to detect arterial air and solid emboli.Methods A pulsed 2 MHz Doppler probe was used to detect the air and clotted whole blood emboli which were introduced into the rabbit aorta via the left renal artery.Results Both the air and the emboli induced similar high intensity transient signals(HITS)in Doppler spectrum.Conclusion Doppler ultrasound can detect the arterial air and solid emboli,but can not distinguish air from solid emboli based on HITS of Doppler spectrum.
出处 《江苏医药》 CAS CSCD 2000年第5期363-364,共2页 Jiangsu Medical Journal
基金 江苏省科委自然基金!资助 (基金编号 :Bk951 31 31 1 )
关键词 多普勒超声 气泡栓子 固体栓子 高强度短暂信号 Doppler ultrasound Air embolus Solid embolus High intensity transient signal
  • 相关文献

参考文献7

  • 11,Spencer MP,Thomas GI,Nicholls SC,et al.Detection of middle cerebral emboli during carotid endarterectomy using transcranial Doppler ultrasonography.Stroke,1990,21:415-423.
  • 22,Russell D.The detection of cerebral emboli using Doppler ultrasound.In:Newell DW and Aaslid R,ed.Transcranial Doppler.New York:Raven Press,1992,207-213.
  • 33,Gibo H,Carver CC,Rhoton AL,et al.Microsurgical anatomy of the middle cerebral artery.J Neurosurg,1981,54:151-169.
  • 44,Russell D,Madden KP,Clark WM,et al.Detection of arterial emboli using Doppler ultrasound in rabbits.Stroke,1991,22:253-258.
  • 55,Ries F,Kaal K,Schultheiss R,et al.Air microbubbles as a contrast medium in transcranial Doppler sonography.J Neuroimag,1991,1:173-178.
  • 66,Consensus Committee of the 9th International Cerebral Hemodynamic Symposium.Basic identification criteria of Doppler microembolic signals.Stroke,1995,26:1123.
  • 77,Babikian VL,Wijman CA,Hyde C,et al.Cerebral microembolism and early recurrent cerebral or retinal ischemic events.Stroke,1997,28:1314-1318.

同被引文献6

  • 1杨芳,李继英.动脉栓塞法复制局灶性脑缺血模型的影响因素探讨[J].中风与神经疾病杂志,2004,21(3):240-242. 被引量:3
  • 2Spencer MP, Thomas GI, Nicholls SC, et al. Detection of middle cerebral artery emboli during carotid endarterectomy using transcranial Doppler ultrasonography. Stroke, 1990,21:415-423.
  • 3Orlandi G, Fanucchi S, Gallerini S, et al. Impaired clearance of microemboli and cerebrovascular symptoms during carotid stenting procedures. Arch Neurol,2005,62 : 1208-1211.
  • 4Georgiadis D, Stets R, Schorch A, et al. Doppler microembolic signals during cardiopulmonary bypass:comparison of two membrane oxygenators. Neurol Res,2004,26:99-102.
  • 5Diehl RR, Samii C, Diehl A. Dynamics and embolic activity of symptomatic intracranial cerebral artery stenosis. Acta Neurol Scand,2002, 106:173-181.
  • 6惠品晶,陈谋森,刘春风,张丽,王之敏,张世明.颈内动脉严重狭窄时脑部症状与Willis环侧支循环建立的关系[J].中风与神经疾病杂志,2002,19(5):278-280. 被引量:34

引证文献1

相关作者