摘要 建立了[18F]-FDG注射液中杂质2-氯-2-脱氧-D葡萄糖(ClDG)含量的HPLC-MS测定方法。色谱柱为Alltima Amino 100A 5μ柱(250×4.6 mm),流动相为5 mmol/L乙酸铵-乙腈(10:90),柱温:35℃;流速:1 mL/min,进样量20μL。ClDG浓度在0.01~0.15 mg/mL范围内与峰面积有良好的线性关系,检出限为0.003 mg/mL,定量限为0.01 mg/mL。首次采用液质联用法检测[18F]-FDG注射液中杂质ClDG含量。 Determination of the concentration of ClDG in [18F]-FDG was established by HPLC-MS.Alltima Amino 100A 5μ column(250 × 4.6 mm) was used with 5 mmol/L ammonium acetate-acetonitrile as mobile phase,at a column temperature of 35 ℃ and a flow rate of 1 mL/min,and injection volume was 20μL.Peak area of ClDG was in a good linear relationship with injection quantity within the range of 0.01-0.15 mg/mL.The detection limite was 0.003 mg/mL and limit of quantification was 0.01 mg/mL.This method is firstly used for the determination of the concentration of ClDG in [18F]-FDG.
出处 《分析试验室》 CAS CSCD 北大核心 2012年第5期93-95,共3页 Chinese Journal of Analysis Laboratory
  • 相关文献

参考文献4

  • 1D. Pavlov, B. O. Myrvold, T. Rogachev, M. Matrakava. J. Power Sources, 85 (2000) 79-91.
  • 2D. P. Boden. J. Power Sources, 107 (2002) 280-300.
  • 3Fludeoxyglucose F 18 injection. Rockville (Md)7 United States Pharmacopeial Convention, 2004:815.
  • 4A. K. Lorenz. Dissertation, Leningrad, Russia, 1953.

共引文献1

同被引文献13

  • 1张悦晗,甄汉深,成莉.蒸发光散射检测器(ELSD)应用概况[J].中华中医药学刊,2007,25(4):831-833. 被引量:33
  • 2Subcommittee on Medical Application of Cyclotron- Produced Radionuclides, Medical Science and Pharmaceutical Committee, Japan Radioisotope Association. Standard for compounds labeled with positron nuclides approved as established techniques for medical use (2001 revision) [ J]. Radioisotopes,2001, 50 (15): 190 - 204.
  • 3Honkomagome, Bunkyo,ku. Description of the "Established Stan- dard Techniques of Labeling Compounds with Positron Nuclides for use as radiopharmaceuticals: Approved by the Subcommittee on Medical Application of Cyclotron- Produced Radionuclides (revision in 2001 )"[J]. Radioisotopes, 2001, 50(5): 1 - 23.
  • 4FDA, HHS. Current Good Manufacturing For Positron Emission Tomography Drugs [ J ]. Federal Register, 2009,74 (236 ) : 65 409.
  • 5Aerts J, Ballinger JR, Behe M, et al. Guidance on current good radio- pharmacy practice for the small-scale preparation of radio- phamaceuticals using automated modules: a European perspective[J]. J Labelled C omp Radiopharm, 2014, 57 (10): 615 -620.
  • 6Natesan R, Van So Le, Hyon - soo H, et al. Guidelines for Good Manufacturing Practices of Radiopharmaceuticals [ M ]. Korea : The Korea Atomic Energy Research Institute, 2001:55 -77.
  • 7Guideline for Quality Assurance and Quality Control of SF - FDG (2 - deoxy - 2 - fluoro - d - glucose) [ EB/OL]. [2011 - 03 - 18 ]. http://www, fnca. mext. go. jp/english/pet/guideline_2, pdf.
  • 8Hung JC. Comparison of Various Requirements of the Quality As- surance Procedures for ISF-FDG Injection [J]. J Nucl Med, 2002,43(11) : 1 495 -1 506.
  • 9Nakao R, Ito T, Yamaguchi M, et al. Simultaneous analysis of FDG, C1DG and Kryptofix 2.2.2 in [ISF] -FDG preparation by high- performance liquid chromatography with UV detection[J]. Nuel Med Bio1,2008,35(2) :239 - 244.
  • 10Dreux M, Lafosse M, Morin - Allory L. The Evaporative Light - Scattering Detector- A Universal Instrument for Non- Volatile Solutes in LC and SFC [J] . LCGC International, 1996,9(3) : 148 - 153.

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部