乳腺癌放射治疗真空垫固定摆位误差分析及优势 被引量:18

Analysis of position set-up tolerance and advantage for breast cancer radiotherapy immobilized by vacuum mold
下载PDF
导出
摘要 背景与目的:目前国内乳腺癌放射治疗使用乳腺托架固定摆位的精度较高,但许多单位没有大孔径CT,因此乳腺托架无法使用。该研究旨在探讨在没有大孔径CT时,乳腺癌术后患者在放射治疗时使用真空垫固定的摆位误差和产生原因。方法:选取2011年11月—2012年2月期间上海交通大学医学院仁济医院放疗科收治的30例乳腺癌术后患者,接受放射治疗并用真空垫固定进行CT定位,使用医科达Precise型加速器上兆伏级电子射野影像系统(electron portal imaging device,EPID)对患者实施锁骨上野和胸壁切线野拍片,并与治疗计划中的射野位置对比验证,统计出X、Y、Z轴3个方向的平均误差及最大误差。结果:X、Y、Z轴平均误差分别为(1.61±1.12)、(1.56±1.12)和(1.50±0.9)mm。结论:乳腺癌术后患者使用真空负压成型垫固定,能保证摆位的精度,在没有固定架条件下实用性高。 Background and purpose: Using the breast bracket would bring patient low position set-up. But the breast bracket is too big to be scanned in normal aperture CT in many radiation centers. This study is to analyze the potision set-up tolerance, with no big aperture CT, for postoperated breast cancer immobilized by vacuum mold when CT scanning during the radiation therapy. Methods: Thirty post-operated breast cancer patients selected from Sep. 2011 to Feb. 2012 were enrolled in this study. EPID was used for taking the posision set-up verification film. Comparing the verification film will obtain the average tolerance and max tolerance. Results: The average tolerance of X, Y and Z axis were (1.61±1.12), (1.56±1.12) and (1.50±0.98)mm. Conclusion: Using the vacuum mold wiU do favor to immobilize the patient position for breast cancer radiotherapy. It would bring patients comfortable position while under radiotherapy and very low tolerance.
出处 《中国癌症杂志》 CAS CSCD 北大核心 2012年第4期283-286,共4页 China Oncology
关键词 乳腺癌 真空垫 电子射野影象系统 摆位误差 Breast cancer Vacuum mold Electron portal imaging device Position set-up tolerance
  • 相关文献

参考文献6

二级参考文献40

  • 1于金明,袁双虎.图像引导放射治疗研究及其发展[J].中华肿瘤杂志,2006,28(2):81-83. 被引量:99
  • 2徐晓娜,吴昊,韩树奎.左侧乳腺癌调强放疗的剂量学研究[J].中华放射肿瘤学杂志,2006,15(3):192-195. 被引量:39
  • 3黄晓波,蒋国樑,陈佳艺,陈兰飞,胡伟刚.乳腺癌调强放射治疗和常规切线野治疗的三维剂量学研究[J].癌症,2006,25(7):855-860. 被引量:62
  • 4David L, Elizabeth J, Adams R, et al. Cardiaea voidance in breast radiotherapy: a comparison of simple shielding techniques with intensity-modulated radiotherapy [ J ], Radiother Oneo1,2001,60 :247-255.
  • 5Larry L, Michael B, Robert C, et al. Intensity modulation to improve dose uniformity with tangential breast radiotherapy : initial clinical experience [ J ]. Int J Radiat Oncol Biol Phys ,2000, 48 : 1559-1568.
  • 6Vicem M,Trank A, Michael B, et al. Significant reductions in heart and long doses using deep insipiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation[J]. Int J Radiat Oncol Biol Phys,2003 ,55 :392-406.
  • 7Bente IGG,Marks LB,Sherous GW,et al.A customized head and neck support system[J].Int J Radiat Oncol Biol Phys,1995,32(1):245-248.
  • 8Marks LB, Yu X, Prosnitz RG, et al. The incidence and functional consequences of RT-associated cardiac perfusion defects, Int J Radiat Oncol Biol Phys,2005,63:214-223.
  • 9Gyenes G, Fornander T, Carlens P, et al. Morbidity of ischemic heart disease in early breast cancer 15 - 20 years after adjuvantradiotherapy. Int J Radiat Oncol Biol Phys,1994,28:1235-1241.
  • 10Gustavsson A, Bendahl P, Cwikiel M, et al. ,No serious late eardiae effects after adjuvant radiotherapy following masteetomy in premenopausal women with early breast eaneer. Int J Radiat Oneol Biol Phys, 1999,43:745-754.

共引文献47

同被引文献142

引证文献18

二级引证文献103

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部