期刊文献+

细菌感染对尿液全自动检测的影响 被引量:2

Effect of bacterial infections on automatic detection of urine
原文传递
导出
摘要 目的观察分析细菌感染对尿液全自动检测的影响。方法应用SysmexUF-1000全自动尿沉渣分析仪及其配套试剂,进行尿红细胞及潜血的测定。结果不同常见种类菌尿对尿红细胞测定结果的影响不同,铜绿假单胞菌影响最大,细菌浓度为104/ml即影响红细胞测定,而细菌浓度为105/ml时对金黄色葡萄球菌及大肠埃希菌有影响;以镜检法为标准,UF-1000检测尿红细胞比较菌尿组假阳性率为30.83%,高于非菌尿组的10.83%,两组差异有统计学意义(P<0.01)。结论 UF-1000尿液分析仪操作简单、快速,有较好的重复性和敏感性,菌尿干扰尿液(沉渣\分析仪对尿细胞、尿潜血)的测定。 OBJECTIVE To observe and analyze the effect of bacterial infections on automatic detection of the urine.METHODS SysmexUF-1000 automated urine sediment analyzer and reagents were taken for the determination of red blood cells and the determination of occult blood in urine.RESULTS Different types of common bacteria had different effects on the determination results of red blood cells,Pseudomonas aeruginosa was influenced most,the concentration of bacteria with 104/ml affected the determination of red blood cells,the concentration of bacteria with 105/ml had effect on Staphylococcus aureus and Escherichia coli;taking microscopic examination as the standard,the false positive rate of red blood cell by UF-1000 Urine Analyzer was 30.83% in bacteriuria group,which was higher than 10.83% in non-bacteriuria group,the difference between the 2 groups was statistically significant(P0.01).CONCLUSION The operation of UF-1000 Urine Analyzer is simple and fast with good reproducibility and sensitivity;bacteriuria can interfere in the determination of urine(sediment/the determination of urine cell by the analyzer,urine occult blood).
出处 《中华医院感染学杂志》 CAS CSCD 北大核心 2012年第9期1986-1987,共2页 Chinese Journal of Nosocomiology
关键词 细菌感染 UF-1000 尿液分析 Bacterial infection UF-1000 Analysis of urine
  • 相关文献

参考文献6

二级参考文献19

  • 1高飞,桂木梅.UF-100和干化学筛检尿路感染的临床意义[J].重庆医学,2004,33(8):1166-1168. 被引量:11
  • 2吴华军,王宇军,陈灵芝.UF-1000尿沉渣流式细胞仪在尿红细胞分析中的应用[J].浙江医学,2009(12). 被引量:5
  • 3毛菊珍,顾国浩,丰斌,沈怡敏.iQ200全自动尿沉渣分析仪假阳性图像的影响因素探讨[J].临床检验杂志,2006,24(5):397-398. 被引量:24
  • 4Srdjatl Stepanovic, Petr Jezek. Isolation of members zhestaphy locoeeus seiuri glvup from urine and their relationship to urinary tract infections. J Clin Micro, 2003, 41:5261.
  • 5Zaman Z, Roggerman S, Verhaegen J. lSnsatifaetory performance of flow eytometer UF - I00 and urine strips in prediction outcome of urine cultures. J Clin Micro, 2001,39:4169.
  • 6李涤生 陈宏础 石自明.临床检验基础[M].北京:人民卫生出版社,1993.188.
  • 7Heang ML. False'hematuria' due to bacteriuria[J]. Arch Pathol Lab Med,1995,119:717 719
  • 8Tsukahara K. Determination of the site responsible for hematuria [J]. Shoni lgaku, 1990,23:773 - 776
  • 9OTTIGER C, SCHAER G, HUBER AR. Time-course of quantitative urinary leukocytes and bacteria counts during antibiotic therapy in women with symptoms of urinary tract infection[J]. Clin Chim Acta, 2007,379(2):36-41.
  • 10KIM SY, KIM YJ, LEE SM. et al. Evaluation of the Sysmex UF-100 urine cell analyzer as a screening test to reduce the need for urine cultures for community-acquired urinary tract infection[J]. Clin Pathol, 2007,128(6) :922-927.