期刊文献+

光纤传输型便携式大面积表面污染监测仪的实验研究 被引量:1

Research on Fiber-based Portable Large Area Surface Contamination Monitor
下载PDF
导出
摘要 以β放射性表面污染测量为对象,基于薄片式大面积塑料闪烁体探测器,建立了光纤传输型大面积塑料闪烁体表面污染监测仪实验装置,探测灵敏面积达1 200 cm2。初步实验表明,在10根光纤作为光收集的条件下,对90Sr-90Yβ面源的探测效率可达到7%。该探测器具有面积大、结构简单、环境适用性强等优点,未来可用于大面积β核素放射性表面污染的测量。 Large area portable surface contamination monitor is suitable for searching the hot point, especially in the situations of large contamination area such as work site radioactive monitoring, nuclear facility decommissioning or nuclear emergency response. This paper focus on beta radioactive surface contamination. Through researching upon the detector which is made of large area plastic scintillation coupled with fibers, the first large area plastic scintillation surface contamination monitor was established. The effective area of detector reaches 1 200 cm2. The verifying experiments indicates that the detection efficiency to 90Sr-90Y plate source comes to 7%. The results show this kind of detector is competent for beta radioactive surface contamination. This work laid a foundation for research and development of the relative instrument.
出处 《辐射防护》 CAS CSCD 北大核心 2012年第3期145-149,共5页 Radiation Protection
关键词 表面污染 实验装置 监测仪 传输型 面积 光纤 便携式 闪烁体探测器 surface contamination monitoring wavelength shifting fibers plastic scintillation nuclear emergency respons
  • 相关文献

参考文献5

  • 1Daisuke Inui, Tadao Hashimoto, Hideyuki Nori- matsu, et al. Development of body surface conta- mination monitor, in The 3rd AOCRP [C]. Tok- oyo, Japan, 2010.
  • 2Maekawa, Sumita, Makino. Thin beta -ray detectors using plastic scintillator combined with wavelength-shifting fibers for surface contamina- tion monitoring [ J]. Energy Society of Japan, 1998, 35.
  • 3Kawagoe K, Sugimoto Y, Takeuchi A, et al. Pe- rformance of preshower and shower-maximum detectors with a lead/plastic-scintillator calorime- ter [J]. Nuclear Instruments and Methods in Phy- sics Research Section A: Accelerators, Spectro- meters, Detectors and Associated Equipment,2002, 487 (3): 275-290.
  • 4Katagiri M, Matsubayashi M, Sakasai K, et al. A compact neutron detector using scintillators with wavelength shifting fibers by backside readout method [ J]. Nuclear Instruments and Methods in Physics Research Section A: Accel-erators, Spectrometers, Detectors and Associated Equipment, 2004, 529 (1-3): 313-316.
  • 5Hou Jie, Luo Zhiping, Ma Jizeng. Optimization for Large area beta surface contamination monitor. In The 3rd AOCRP [C]. Tokoyo, Japan, 2010.

同被引文献4

  • 1INUI D, et al. Development of body surface contamination monitor [ C ]// The 3rd AOCRP. Tokoyo, Japan. 2010.
  • 2Bukin DA, Druzhinin VP, Golubev VB. Scintillation counter with WLS fiber readout [ J ]. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997,384 ( 2 -3): 360-363.
  • 3中国科学院数学与系统科学研究院,中国标准化研究院,北京师范大学,等.GB/T3358.1-2009/ISO3534-1:2006统计学词汇及符号第1部分:一般统计术语与用于概率的术语[S].北京:中国标准出版社,2009:24.
  • 4中国原子能科学研究院.GB/T8997--2008d、B表面污染测量仪与监测仪的校准[S].北京:中国标准出版社.2008:2.

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题