冲洗和扫描条件对胶片剂量响应的影响

Analysis the effects of fill dosimetry during processing and scanning
原文传递
导出
摘要 目的分析冲洗和扫描条件对胶片剂量与吸光度(A)值的影响,建立基于卤化银胶片剂量仪的临床质量保证程序。方法选择不同洗片温度、不同批次胶片、不同批次药水及同批次药水和胶片分析剂量与A值的响应。测试扫描仪不均匀性、稳定工作时间和噪声对胶片剂量响应的影响。结果同一剂量水平A值随温度升高而增大;不同批次胶片剂量与A值响应差异较大;不同批次药水、胶片剂量与A值响应差异明显;同一批次胶片剂量与A值响应与药水时间呈现一定规律。扫描仪扫描的不均匀性最大达0.03,相同剂量A值在预热10min后稳定,噪声影响随剂量和分辨率增加而加大。结论最佳洗片温度为29~31℃;不同批次胶片不能混用,更换胶片、药水时需重新获取刻度片。扫描仪需进行不均匀性校准,最佳扫描设置为72dpi、16bit,且要预热5—10min。 Objective To explore the effects of different flushing and scanning processing on optical density (A) responding and set up the clinic quality assurance protocol based on silver halide emulsions radiographic films. Methods Setting different flushing temperature and choosing different batch's film and developer and fixer or fixer in the same batch with different analyzing dose were performed to analyze the effect on A value;The effect of light uniform, the stabilize work time and the noise of different scanning resolution were discussed. Results The A value at the same dose level would enhance as the temperature increased; the responding curve of dose and A value with different batches of films varied a lot; the responding curve of dose and A value with different batches of films and developer and fixer solution had marked variations;the responding of dose and A with the same batches would show some low. The heterogeneity of the scanner would achieve 0. 03 ; the A value of the same dose would gradually steady while the time of scanning was more than 10 minutes; the affect of noise would increase as the dose and resolution ratio increased. Conclusions The best processing temperature is 29 -31~C. Different batches of film couldn't be confounded. A new calibration must be obtained when the film dosimetry to evaluate dose distribution is used. A 5 - 10 min warm up for stabilize work and the best setting resolution/depth lever are 72 dpi/16 bit for scanning films are determined.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2012年第3期285-288,共4页 Chinese Journal of Radiation Oncology
关键词 剂量学 胶片 吸光度 冲洗与扫描条件 Dosimetry, film Optical density Flushing and scanning condition
  • 相关文献

参考文献7

  • 1Pai S, Das IJ, Dempsey JF, et ah TG-69:radiographic film for megavoltage beam dosimetry. Med Phys ,2007,34:2228-2258.
  • 2Childress NL, Rosen II. Effect of processing time delay on the dose response of Kodak EDR2 film. Med Phys, 2004, 31: 2284- 2288.
  • 3Childress NL, Dong L, Rosen I1. Rapid radiographic film calibration for IMRT verification using automated MLC fields. Med Phys, 2002,29 : 2384 -2390.
  • 4李勤,李光俊,韩军,孙长庚.调强验证中胶片刻度方法的研究[J].中华放射肿瘤学杂志,2007,16(5):384-385. 被引量:6
  • 5Mersseman B, De Wagter C. Characteristics of a commereially available film digitizer and their significance for film dosimetry. Phy Med Biol, 1998,3 : 1803-1812.
  • 6Wilcox E, Daskalov G, Nedialkova L. Comparison of the Epson Expression 1680 flatbed and the Vidar VXR-16 Dosimetry PROTM film scanners for use in IMRT dosimetry using Gafchromic and radiographic film. Med Phys,2007 ,54 :41-48.
  • 7Baruffaldi F, Angelini AL, Testi D, et al. CCD film digitizers in clinical practice: evaluation of the main properties. Med Inform Internet Med,2001,26 : 101-114.

二级参考文献1

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部