两种不同制瓣方式对LASIK术角膜瓣厚度的影响 被引量:5

Comparison of LASIK flap thickness created with Ziemer LDV femtosecond laser and Moria M2 mechanical microkeratome
下载PDF
导出
摘要 目的:比较ZiemerLDV飞秒激光与MoriaM2机械板层刀不同制瓣方式对LASIK术角膜瓣厚度的影响。方法:选取近视患者100例200眼(近视度数-2.00~-12.00D,散光度数0.00~-3.50D),分别应用ZiemerLDV飞秒激光(设定角膜瓣切削厚度110μm)与MoriaM2(110刀头)微型角膜刀制作角膜瓣,各50例100眼。术后1wk,应用眼前节RTVueFD-OCT测量,每个角膜上0°,45°,90°,135°四条子午线所在截面上中心点、距离中心1,2,3mm特定7个点(共计28个点)的角膜瓣厚度测量,并作分析比较。结果:角膜瓣中央点厚度:ZiemerLDV飞秒激光组108.69±11.75μm,MoriaM2微型角膜刀组130.75±13.36μm,所有观测点(共28点)两组间有明显统计学差异(P<0.01)。角膜中心点厚度与预计值差值比较:ZiemerLDV飞秒激光组9.70±6.84μm,MoriaM2微型角膜刀组21.63±11.79μm。ZiemerLDV飞秒激光组预计值差值明显低于MoriaM2微型角膜刀组(t=17.493,P<0.01)。ZiemerLDV飞秒激光组与MoriaM2微型角膜刀组同一截面中各点角膜瓣厚度值差异有统计学差异(P<0.01)。结论:角膜瓣制作方式相比,在角膜瓣的精确性、可预测性等方面ZiemerLDV飞秒激光优于MoriaM2微型角膜刀。ZiemerLDV组与MoriaM2机械板层刀组都未表现角膜瓣厚度的均匀性。 AIM: To compare corneal flap thickness created with Ziemer LDV femtosecond laser and Moria M2 mechanical microkeratome. METHODS: A retrospective analysis of 100 cases (200 eyes) were performed with myopia and astigmatism (sphere -2.00- -12.00 diopters[D];cylinder 0.00- -3.50 diopters[D]). Fifty cases (100 eyes) were treated with Ziemer LDV femtosecond laser (Intended flap thickness was 110μm) and fifty cases (100 eyes) were treated with Moria M2 mechanical microkeratome using the 110μm head. Corneal flaps parameters were measured at center and 1mm,2mm,3mm from center located in meridian 0°,45°,90°,135°section, each section including seven points evaluated by RTVue FD-OCT one week after procedure. All measurement points were analyzed. RESULTS: The mean flap thickness at center was 108.69±11.75μm in Ziemer group and 130.75±13.36μm in Moria M2 group. There was a statistical difference in relative points between the two groups (t=24.791,P0.01). The mean value of each section by subtraction a target flap thickness of 110μm was 9.70±6.84μm in Ziemer group and 21.63±11.79μm in Moria M2 group at center. Ziemer group was significantly less than Moria M2 group (t=17.493,P0.01). Flap thickness of seven points in the same section were significantly different in both Ziemer group and Moria M2 group(P0.01). CONCLUSION:Flap creation with Ziemer LDV femtosecond laser was superior to Moria M2 mechanical microkeratome in precision and predictability. The flap thickness of different point in the same section was not equal in two groups.
出处 《国际眼科杂志》 CAS 2012年第6期1056-1058,共3页 International Eye Science
关键词 Ziemer LDV飞秒激光 MORIA M2机械板层刀 角膜瓣 厚度 Ziemer LDV femtosecond laser Moria M2 mechanical microkeratome corneal flap thickness
  • 相关文献

参考文献9

  • 1Li H,Sun T, Wang M, et al. Safety and effectiveness of thin-flap LASIK using a femtosecond laser and microkeratome in the correction of high myopia in Chinese patients. J Refract Surg 2010;26(2):99-106.
  • 2Sutton G, Hodge C. Accuracy and precision of LASIK flap thickness using the IntraLase femtosecond laser in 1000 consecutive cases. J Refract Surg 2008;24(8):802-806.
  • 3Li Y,Netto MV, Shekhar R, et al. A longitudinal study of LASIK flap and stromal thickness with high-speed optical coherence tomography. Ophthalmology 2007;114(6):1124-1132.
  • 4Kim JH, Lee D, Rhee KI. Flap thickness reproducibility in laser in situ keratomileusis with a femtosecond laser: optical coherence tomography measurement. J Cataract Refract Surg 2008 ;34(1):132-136.
  • 5Kezirian GM, Stonecipher KG. Comparison of the Intralase femtosecond laser and mechanical keratomes for laser in situ keratomileusis. J Cataract Refract Surg 2004;30(4):804-811.
  • 6Shemesh G, Dotan G, Lipshitz I. Predictability of corneal flap thickness in laser in situ keratomileusis using three different microkeratomes. J Refract Surg 2002;18(3Suppl):s347-351.
  • 7RosasSalaroli CH, Li Y, Zhang X, et al. Repeatability of laser in situ keratomileusis flap thickness measurement by Fourier-domain optical coherence tomography. J Cataract Refract Surg 2011;7(4):649-654.
  • 8Ou RJ,Shaw EL,Glasgow BJ. Keratoectasia after laser in situ keratomileusis (LASIK):evaluation of the calculated residual stromal bed thickness. Am J Ophthalmol 2002;134(5):771-773.
  • 9于志强,许烨,姚佩君,秦冰,周行涛,褚仁远.准分子激光手术不同制瓣方式角膜瓣厚度的研究[J].中华眼科杂志,2010,46(3):203-208. 被引量:23

二级参考文献19

  • 1Izquierdo L Jr, Henriquez MA, Zakrzewski PA. Detection of an abnormally thick LASIK flap with anterior segment OCT imaging prior to planned LASIK retreatment surgery. J Refract Surg, 2008, 24 : 197-199.
  • 2Muallem MS, Yoo SH, Romano AC, et al. Flap and stromal bed thickness in laser in situ keratomileusis enhancement. J Cataract Refract Surg,2004 ,30 :2295-2302.
  • 3Li Y, Shekhar R, Huang D. Corneal pachymetry mapping with high-speed optical coherence tomography. Ophthalmology, 2006, 113 : 792-799.
  • 4Li Y, Netto MV, Shekhar R, et al. A longitudinal study of LASIK flap and stromal thickness with high-speed optical coherence tomography. Ophthalmology ,2007, 114 : 1124-1132.
  • 5Cheng AC, Rao SK, Lau S, et al. Central corneal thickness measurements by ultrasound, Orbscan II , and Visante OCT after LASIK for myopia. J Refract Surg, 2008,24:361-365.
  • 6Thomas J, Wang J, Rollins AM, et al. Comparison of corneal thickness measured with optical coherence tomography, ultrasonic pachymetry, and a scanning slit method. J Refract Surg, 2006, 22:671-678.
  • 7Cheng AC, Ho T, Lau S, et al. Measurement of LASIK flap thickness with anterior segment optical coherence tomography. J Refract Surg, 2008, 24:879-884.
  • 8Kim JH, Lee D, Rhee KI. Flap thickness reproducibility in laser in situ keratomileusis with a femtosecond laser: optical coherence tomography measurement. J Cataract Refract Surg, 2008,34 : 132- 136.
  • 9Wang J, Thomas J, Cox I, et al. Noncontact measurements of central corneal epithelial and flap thickness after laser in situ keratomileusis. Invest Ophthalmol Vis Sci,2004,45 : 1812-1816.
  • 10Shemesh G, Dotan G, Lipshitz I. Predictability of corneal flap thickness in laser in situ keratomileusis using three different microkeratomes. J Refract Surg, 2002,18:347-351.

共引文献22

同被引文献47

  • 1何飞,程亚.飞秒激光微加工:激光精密加工领域的新前沿[J].中国激光,2007,34(5):595-622. 被引量:95
  • 2Li H, Sun T, Wang M,et al. Safety and effectiveness of thin-flap LASIK using a femtosecond laser and microkeratome in the correction of high myopia in Chinese patients. J Refract Surg 2010 ,26 ( 2 ) :99-106.
  • 3Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 2005 ,31 (1) : 156-162.
  • 4Ambrosio R Jr, Dawson DG, Salomao M,et al. Corneal ectasia after LASIK despite low preoperative risk: tomographic and biomechanical findings in the unoperated, stable, fellow eye. J Refract Surg 2010,26 (11) :906-911.
  • 5Shin J, Kim TW, Park SJ,et al. Changes in biomechanical properties of the cornea and intraocular pressure after myopic laser in situ keratomileusis using a femtosecond laser for flap creation determined using ocular response analyzer and goldmann applanation tonometx'y. J Glaucoma 2013. Epub 2013 Jun 25.
  • 6Troost R,Vogel A, Beck S,et al. Clinical comparison of two intraocular pressure measurement methods: SmartLens dynamic observing tonography versus Goldmann. C, raefe,s Arch Clin Exp Ophthalmol 2001 ,239(12) :889-892.
  • 7Chen S, Chen D, Wang J,et al. Changes in ocular response analyzer parameters after LASIK. J Refract Surg 2010 , 26 ( 4 ) : 279- 288.
  • 8Medeiros FW, Sinha-Roy A, Alves MR,et al. Biomechanical corneal changes induced by different flap thickness created by femtosecond laser. Clinics ( Sao Paulo ) 2011 ,66(6) :1067-1071.
  • 9Ou RJ, Shaw EL, Glasgow BJ. Keratoectasia after laser in situ keratomileusis (LASIK) :evaluation of the calculated residual stromal bed thiekness. Am J Ophthalmol 2002 , 134 ( 5 ) :771-773.
  • 10Hamilton DR,Johnson RD, Lee Bourla N. Differences in the corneal biomechanical effects of surface ablation compared with laser in situ keratomileusis using a microkeratome or femtosecond laser. J Cataract Refract Surg 2008 , 34 ( 12 ) : 2049- 2056.

引证文献5

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部