两种不同检测系统测定血清总蛋白的结果比对与偏倚评估 被引量:3

Results comparison and bias estimation between two different biochemical detection systems for serum total protein assays
下载PDF
导出
摘要 目的探讨本实验室内待测系统与参比系统测定血清总蛋白(TP)的可比性。方法参照临床和实验室标准化协会(CLSI)颁发的EP9-A2文件,每天取患者样本8份,分别用待测系统与参比系统测定TP,并且每次均进行平行管测定,共测定5天;记录试验数据,经离群点检验后,计算相关系数和回归方程,进行偏倚评估。结果测定结果无方法内离群点及方法间离群点;两种检测系统TP测定结果的相关系数和回归方程分别为r=0.999(P=0.000),Y=1.035X-1.211(X、Y分别为参比系统、待测系统的TP测定结果);两种检测系统在TP医学决定水平(Xc)处的预期偏倚(Bc)小于可接受偏倚(P<0.05)。结论上述两种检测系统测定血清TP的结果基本一致,在本实验室内可以使用两种系统检测血清TP。 Objective To investigate the comparability of serum total protein (TP) test results between our detection system and the reference system. Methods According to the EP9-A2 document described by the Clinical and Laboratory Standards Institute (CLSI), 40 patient samples were analyzed in 5 days, and each pa- tient sample was analyzed in duplicate using both our detection system and the reference system. After the out-lier tests, the correlation coefficient, regression equation and bias between the two detection systems Were cal-culated. Results There were no within-method outlier or between-method outlier in the experimental data. The correlation coefficient and the regression equation for serum TP between two detection systems were 0.999 (P=0.000)and Y=1.035X-1.211, respectively. The predicted bias at each given medical decision level was less than the acceptable bias (P〈0.05). Conclusion The serum TP test values from our detection system agree with the results of reference system, and each of them can be used to measure serum TP in our laboratory.
出处 《实验与检验医学》 CAS 2012年第2期144-146,共3页 Experimental and Laboratory Medicine
关键词 检测系统 总蛋白 偏倚 Detection System Total protein Bias
  • 相关文献

参考文献4

二级参考文献3

  • 1National Committee for Clinical Laboratory Standards.Method comparison and bias estimation using patient samples,approved guideline,EP9-2A,Pennsylvania.NCCLS,1986:1-14.
  • 2Chen Y,WuW,Li LJ,et al.Comparison of the result for three automated immunoassay system in determining serum HBV markers.Clin Chim Acta,2006,372:129-133.
  • 3邱玲,程歆琦,刘荔,刘福成,刘茜,葛增梅.多台生化分析仪多项目同时进行比对的实验研究设计及应用[J].中华检验医学杂志,2007,30(9):1001-1004. 被引量:55

共引文献110

同被引文献16

  • 1吕赛平,伍志杰,邹学森.自动生化分析仪精密度性能的评价[J].现代检验医学杂志,2007,22(4):48-49. 被引量:12
  • 2Clinical and Laboratory standards Institute Method Comparison and Bias Estimation Using Patient Samples[ S]. EP9 - A2, CLSI. 2002.
  • 3Smolcic VS, Bilic-Zulle L, Fisic E. Validation of methods perfor- mance for routine biochemistry analytes at Cobas 6000 analyzer series module c501 [J]. Biochem Med (Zagreb), 2011,21 (2):182- 190.
  • 4Clinical and Laboratory Standards Institute. Evaluation of Precision Performance of Quantitative Measurement Methods [S]. EPS-A2, CLS], 2004.
  • 5De Koninck AS, De Decker K, Van Bocxlaer J, et al. Analytical performance evaluation of four cartridge-type blood gas analyzers [J]. Clin Chem Lab Med, 2012, 50(6):1083-1091.
  • 6Lebitasy MP, Latinne D, Maisin D, et al. Clauss assay and f/brino- gen protein estimated by capillary zone electrophoresis [J]. Clin Chem Lab Med, 2011, 49(4):689-693.
  • 7Carlier P, Harlka N, Bailly R, et al. Laboratory evaluation of the new Access cytomegalovirus immunoglobulin IgM and IgG assays [J]. J Clin Virol, 2010, 49(3):192-197.
  • 8Kos S, van Meerkerk A, van der Linden J, et al. Validation of a new Generation POCT glucose device with emphasis on aspects important for glycemie control in the hospital care[J]. Clin Chem Lab Med,2012,50(9) :1573-1580.
  • 9CLSI. EP9-A2 Method comparison and bias estimation using patient samples[S]. Wayne, PA: CLSI, 2002.
  • 10黄龙,刘志伟,郑定容.血清和肝素钠抗凝血浆电解质测定浓度的比较[J].实用医技杂志,2007,14(33):4547-4548. 被引量:5

引证文献3

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部