摘要 目的:探讨三套检测系统对葡萄糖检测结果的一致性。方法:以TBA-120FR全自动生化分析仪为参考方法,以VITROS350干化学分析仪和BIOSEN C-line葡萄糖乳酸分析仪为评价方法,对患者葡萄糖进行检测,用配对t检验及回归分析评价检测结果的相关性。结果:三种检测系统葡萄糖检测结果进行配对t检验,葡萄糖测定结果比较差异无统计学意义(P>0.05),相关系数分别为0.996、0.993、0.994。结论:通过比较分析证明实验室内不同检测系统应定期进行对比,为保证检测结果的准确性与一致性提供客观依据。
作者 白文言
出处 《吉林医学》 CAS 2012年第11期2398-2398,共1页 Jilin Medical Journal
  • 相关文献

参考文献4

二级参考文献4

共引文献17

同被引文献28

  • 1贾伟平.中国动态血糖监测临床应用指南(2012年版)[J].慢性病学杂志,2013(5):321-330. 被引量:12
  • 2Tendl KA, Christoph J, Bohn A, et al. Two site evaluation of the performance of a new Generation point-of-care glu- cose meter for use in a neonatal intensive care unitE J2. Clin Chem Lab Med,2013,51 (9) :1747-1754.
  • 3Soares Dde S,Brando RR, Mouro MR,et al. Relevance of routine testing in low risk patients undergoing minor and medium surgical procedures E J2. Rev Bras Anestesiol, 2013,63 (2) :197-201.
  • 4Buhl M, Bosnjak E, Vendelbo MH, et al. Direct effects of locally administered lipopolysaccharide on glucose, lipid, and protein metabolism in the placebo-controlled, bilater- ally infused human leg[J]. J Clin Endocrinol Metab, 2013, 98(5) ;2090-2099.
  • 5Pftitzner A, Schipper C, Ramljak S, et al. Determination of hematocrit inter[erence in blood samples derived Jrom pa tients with different blood glucose concentrations [J]. J Diabetes Sci Technol,2013,7(1) : 170-178.
  • 6Lucarelli F,Ricci F,Caprio F,et al. GlucoMen day contin- uous glucose monitoring system., a screening for enzymatic and electrochemical interferentsJ. J Diabetes Sci Techn- ol,2012,6(5) :1172-1181.
  • 7Yu X,Limin W,Jiang H,et al. Prevalence and Control of Diabetes in Chinese Adults [J]. JAMA, 2013,310(9):948- 958.
  • 8Ho DP,Kyoung JL,Hyung RY,et al. Design of a portable urine glucose monitoring system for health care [J]. Computers in Biology and Medicine,2005,35(4):275-286.
  • 9Liu X,Zhang S,Tan P,et al. A plasmonic blood glucose monitor based on enzymatic etching of gold nanorods [J]. Chemical Communications ,2013,49(18) : 1856-1858.
  • 10van Dijk TH,Laskewitz AJ,Grefhorst A,et al. A novel approach to monitor glucose metabolism using stable isotopically labelled glucose in longitudinal studies in mice [J]. Laboratory Animals,2013,47(2):79-88.

引证文献5

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部