Developments of a 2D position sensitive neutron detector 被引量:2

Developments of a 2D position sensitive neutron detector
原文传递
导出
摘要 China Spallation Neutron Source (CSNS), one project of the 12th Five-Year-Plan scheme of China, is under construction in Guangdong province. Three neutron spectrometers will he installed during the first phase of the project, and two-dimensional position sensitive thermal neutron detectors are required. Before the construction of the neutron detectors, a prototype of a two-dimensional 200 mm^200 mm Multi-wire Proportional Chamber (MWPC) with Ar/CO2 (90/10) flowing gas has been constructed. In 2009, the prototype was tested with the 55Fe X-ray using part of the electronics, and performed well. Following the test in 2009, the neutron detector was constructed with the complete electronics and filled with the 6 atm. 3He^2.5 atm. C3Hs gas mixture in 2010. The neutron detector has been primarily tested with an Am-Be source. In this paper, new developments of the neutron detector including the design of the high pressure chamber, the optimization of the gas purifying system and the gas filling process will be reported. The results and discussion are also presented in this paper. China Spallation Neutron Source (CSNS), one project of the 12th Five-Year-Plan scheme of China, is under construction in Guangdong province. Three neutron spectrometers will he installed during the first phase of the project, and two-dimensional position sensitive thermal neutron detectors are required. Before the construction of the neutron detectors, a prototype of a two-dimensional 200 mm^200 mm Multi-wire Proportional Chamber (MWPC) with Ar/CO2 (90/10) flowing gas has been constructed. In 2009, the prototype was tested with the 55Fe X-ray using part of the electronics, and performed well. Following the test in 2009, the neutron detector was constructed with the complete electronics and filled with the 6 atm. 3He^2.5 atm. C3Hs gas mixture in 2010. The neutron detector has been primarily tested with an Am-Be source. In this paper, new developments of the neutron detector including the design of the high pressure chamber, the optimization of the gas purifying system and the gas filling process will be reported. The results and discussion are also presented in this paper.
出处 《Chinese Physics C》 SCIE CAS CSCD 2012年第6期525-530,共6页 中国物理C(英文版)
基金 Supported by National Natural Science Foundation of China (11127508)
关键词 thermal neutron detector two dimensional MWPC Am-Be neutron source thermal neutron detector, two dimensional MWPC, Am-Be neutron source
  • 相关文献

参考文献9

  • 1Knott R B, Smith G C, Watt Get al. Nuclear Instruments and Methods in Physics Research A, 1997, 392:62.
  • 2YU B, Harder J A, Mead J A et al. Nuclear Instruments and Methods in Physics Research A, 2003, 513:362.
  • 3SHAIKH A M, DESAI S S, PATRA A K. Pramana - J. Phys., 2004, 63:465.
  • 4BOIE R A, FISCHER J, INAGAKI Y et al. Nuclear Instruments and Methods, 1982, 200:533.
  • 5WANG Xiao-Hu, ZHU Qi-Ming, CHEN Yuan-Bo et al. Chinese Physics C (HEP & NP), 2008, 32:903.
  • 6Radeka V, Schaknowski N A, Smith G C et al. Nuclear Instruments and Methods in Physics Research A, 1998, 419: 652.
  • 7XIE Yi-Gang et al. Particle Detector and Data Acquisition. First Edition. Beijing: Sciece Press, 2003. 73.
  • 8Veenhof R. Garfield Help Pages. http://garfield.web.cern. ch/garfield/.
  • 9Ziegler J F. The Stopping and Range of Ions in Matter, vol. 2-6, Pergamon Press, 1977-1985. http://www.srim.org.