摘要
建立了测定食品添加剂磷酸中砷的原子荧光光谱法,测试中砷的相对标准偏差最高为3.07%,回收率在98.7%~101.2%之间、检出限为0.09μg/L;本法相对GB3149-2004《食品添加剂磷酸》国家标准中砷的测定方法而言操作简便、快速、准确、灵敏度高,值得推广。
This paper established the determination of arsenic in food additive phosphoric acid by Atomic fluorescence spectrometry, and the maximum of the relative standard deviation in arsenic test is 3.07%, the recovery rate was 98.7% - 101.2%, the detection limit is 0.09 μg/L; this method is easy operation, rapid, accurate, high sensitivity, worth popularizing in terms of relatively GB3149 -2004 《food additive phosphoric acid》 national standard method for the determination of arsenic.
出处
《山东化工》
CAS
2012年第6期53-55,58,共4页
Shandong Chemical Industry
关键词
原子荧光光谱法
食品添加剂磷酸
砷
atomic fluorescence spectrometry
food additive phosphoric acid
arsenic