摘要 目的对贝克曼库尔特LH750全自动血细胞分析仪的应用性能进行评价。方法采用EDTA-K。抗凝静脉血液标本,对LH750血细胞分析仪的准确度、精密度、携带污染率、线性范围比较试验进行检测,将结果与贝克曼库尔特公司提供的设计标准比较,应用两种不同进样模式检测标本,比较两者之间的差异。结果LH750全自动血细胞分析仪的准确度、精密度、携带污染率、线性范围等结果,均符合设计规定要求,两者进样模式在允许偏倚范围之内。结论LH750全自动血细胞分析仪应用性能良好,能满足血液常规分析的一般要求。 Objective To evaluate the application performance of Beckman Coulter LH750 automated hematology ana- lyzer. Methods Using of EDTA K2 anticoagulant venous blood samples to detect accuracy, precision, contamina- tion carrying over rate, linear range, and compared them with the designed standard provided by Beckman Coulter company. Samples were detected with two different sampling modes and the results were compared. Results The accuracy, precision, contamination carrying over rate, linear range of Beckman Coulter LH750 automated hematolo- gy analyzer accord with the designed requirement of Beckman Coulter company. Biases of two different sampling modes were in the permitted range. Conclusion The Beckman Coulter LH750 automated hematology analyzer is well performed and achieves the clinical demand of hematology analysis.
作者 李继恩
出处 《中国误诊学杂志》 CAS 2012年第16期4173-4175,共3页 Chinese Journal of Misdiagnostics
关键词 血细胞计数/仪器和设备 Blood Cell Count/instrumentation
  • 相关文献

参考文献4

二级参考文献23

  • 1丛玉隆.血液分析仪应用中的几个问题[J].中华医学检验杂志,1994,17(6):325-326. 被引量:68
  • 2Kasoff J. In the Foreseeable Future. Advance for Administrtors of the Laboatory, 2007,16: 52.
  • 3Smith TJ. Automating the hematology Lab. Advance for Administrtors of the Laboatory, 2008,17 : 68-70.
  • 4Molero T, Lemes A, Iglesia S, et al. Monoclonal antibody fluorescence for routine lymphocyte subpopulation analysis with the Abbott CELL-DYN Sapphire hematology analyzer. Int J lab Hematol, 2007,29:446-453.
  • 5Bagdasaryan R, Zhou Z, Tierno B, et al. Neutrophil VCS parameters are superior indicators for acute infection. Lab Hematol, 2007,13 : 12-16.
  • 6Chaves F, Tiemo B, Xu DS. Quantitative determination of neutrophil VCS parameters by Coulter automated hematology analyzer: new and reliable indicators for acute bacterial infection. American Journal of Clinical Pathology, 2005,124 : 440-444.
  • 7Chaves F,Tierno B, Xu DS. Neutrophil volume distribution width (NDW) : A new, automated hematologic parameter for acute infection. Archives of Pathology & Laboratory Medicine, 2006, 130:378-380.
  • 8Tiemo B, Chaves F, Xu DS. The potency of prevalence. Am J Clin Pathol, 2006,125: 475-476.
  • 9Carol Briggs, Anabela Da Costa, Lyn Freeman, et al. Development of an automated malaria discrlminant factor using VCS technology. Am J Clin Pathol, 2006,126:691-698.
  • 10Fourcade C, Casbas M, Belaouni H, et al. Automated detection of malaria by means of the hematology analyzer Coulter GEN. S. Clinical & Lab Hematol, 2004,26:367-372.