TECAN EVO Clinical全自动样品处理系统的性能评价

Evaluation and acception the capability of the Pre-baking system
原文传递
导出
摘要 目的评价与验收TECANEVO Clinical全自动样品处理系统的性能。方法用比重为1.000的蒸馏水作为测试品,分别做进行10μl和100μl两个量程的加样实验,对TECANEVOClinical全自动样品处理系统自动加样器的精密度、准确度进行测试性能评价。选取健康体检标本24份,乙肝患者标本48份,HBsAg弱阳性携带者标本24份进行携带污染率测试。结果8通道加样针的精密度均小于3.5%、0.75%。结论准确度、精密度及携带污染率均在可接受范围内,TECANEVOClinical全自动样品处理系统的性能可靠,能够满足临床实验的需要。 Objective To evaluate and accept the capability of the full autokinetic system about TECAN EVO Clinical. Methods Use distilled water with specific gravity is 1. 000 for testing accuracy and precision about the pre-bak- ing system. And use 24 numbers of healthy specimen ,48 numbers of B liver sufferet's specimen and 24 numbers of tahe the B liver virus's specimen for testing the rate of taking a stain. Result ALL of the instrument index were credible. Conclusion Accuracy , precision and the rate of taking a pollution are in extension, the capability ofthe Pre- baking system about TECAN EVO is acceptable so that it can settle for the clinical need.
出处 《中国误诊学杂志》 CAS 2012年第17期4478-4479,共2页 Chinese Journal of Misdiagnostics
关键词 实验室技术和方法/仪器和设备 Laboratory Techniques and Procedures/instrumentation