摘要 通过检测透皮抽取出的人体细胞间液中葡萄糖浓度来预测血糖浓度是一种重要的微创血糖检测技术。但由于抽取出的细胞间液量很少,需要进行稀释以便输送和收集,然后根据稀释后的体积来计算稀释前的细胞间液浓度。设计和加工了一种微流量计用以测量稀释后细胞间液的体积,并搭建了性能测试实验平台,测试了流量计测量结果的重复性和线性相关性。实验表明:流量计测量范围为4~16μL;对15μL的样品10次测量的变异系数为6.2%;13个不同体积样品的测量结果相关系数为0.988。微流量计能够满足测量要求。 It is an important minimally invasive blood glucose detection technique that predicting the blood glucose level by glucose concentration of the extracted interstitial fluid. However, the extracted interstitial fluid should be diluted to deliver and collect because of its small amount. Then the concentration before diluted can be calculated by the volume measured after diluted. A kind of micro flowmeter is designed and fabricated to measure the volume after diluted, a performance test experimental platform is built up to test the repeatability and linear correlation. Experimental results show that the measurement range is 4 - 16 μL and the coefficient of variation of 10 times measurement for 15 μL samples is 6.2 % and the correlation coefficient of 13 samples with different volumes is 0. 988. The measurement range of micro-flowmeter is 4 - 16 μL which can meet the requirements of measurement.
出处 《传感器与微系统》 CSCD 北大核心 2012年第7期117-119,125,共4页 Transducer and Microsystem Technologies
基金 国家自然科学基金资助项目(308002389 61176107)
关键词 微创血糖检测 微流量计 细胞间液透皮抽取 细胞间液体积 minimally invasive blood glucose detection micro flowmeter interstitial fluid transdermal extraction volume of the extracted interstitial fluid
  • 相关文献

参考文献9

  • 1姜利英,任丽杰,陈青华,张法全,崔光照.血糖检测技术研究进展[J].传感器世界,2009,15(10):6-10. 被引量:8
  • 2Mitragotri S. Synergistic effect of enhancers for transdermal drug delivery[ J]. Pharmaceutical Research ,2000 ( 11 ) :1354 -1359.
  • 3Luis J, Park E J, Meyer Jr R J. Rectangular cymbal arrays for improved ultrasonic transdermal insulin delivery [ J 1. Acoust Soc Am,2007,122 (4) :2022 -2030.
  • 4于海霞.关于组织液抽取和血糖连续检测的研究[D].天津:天津大学,2006.
  • 5Loo Y L, Willett R L, Baldwin K W, et al. Additive, nanoscale patterning of metal films with a stamp and a surface chemistry mediated transfer process : Applications in plastic electronics [ J ]. Applied Physics Letters ,2002,81 (3) :562 -564.
  • 6Nakamura M,Kato S, Aoki T, et al. Role of terminal OH groups on the electrical and hydrophilic properties of hydro-oxygenated amorphous TiOx : OH thin films[ J ]. Journal of Applied Physics, 2001,90 (7) :3391 -3395.
  • 7Nakamura M, Aoki T, Hatanaka Y. Hydrophilic characteristics ofRF-sputtered amorphous TiO2 film [ J ]. Vacuum, 2000,59 ( 2 - 3 ) :506 -513.
  • 8Schueller 0 J A, Duffy D C, Rogers J A, et al. Reconfigurable dif- fraction gratings based on elastomeric microfluidic devices [ J ]. Sensors and Actuators A, 1999,78 (2 -3 ) : 149 -159.
  • 9Yeh T T,Espina P I. An intelligent ultrasonic flow meter for improved flow measurement and flow calibration facility [ C ]//Proc of IEEE Instrumentation and Measurement Technology Conference ,2001 : 1741 -1746.

二级参考文献16

共引文献7

同被引文献6

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部