目的:探讨羊水板层小体计数预测胎肺成熟度的临床应用价值。方法:选择156例单胎孕妇作为研究对象,孕28~38周,每份羊水分3份分别进行LBC计数、L/S比值测定及泡沫试验。结果:羊水LBC与L/S比值成正线性相关,相关系数(r)为0.623,有极显著统计学意义(P<0.001);LBC与泡沫试验呈等级相关,相关系数为(r)为0.781,有极显著统计学意义(P<0.001)。以LBC≤10×109/L作为判断胎儿肺不成熟的临界值,以LBC≥35×109/L作为判断胎儿肺成熟的临界值,比较羊水LBC计数各预测值率与L/S比值差异无统计学意义(P>0.05);与泡沫试验相比,羊水LBC计数的各预测值明显高于泡沫试验,差异有统计学意义(P<0.01)。结论:羊水LBC计数在预测胎肺成熟度方面优于传统的泡沫试验,较L/S比值省时、省力,是临床中用于判断胎儿肺成熟方法的首选。
Objective:To explore the clinical application value of lamellar body count in amniotic fluid for prediction of fetal lung maturity.Methods:A total of 156 single pregnant women were selected as study objects,the amniotic fluid samples underwent LBC count,L/S detection,and foam test simultaneously during 28-38 gestational weeks.Results:There was a positive linear correlation between LBC count and L/S ratio(γ=0.623,P〈0.001);there was a rank correlation between LBC count and foam test(γ=0.781,P〈0.001);LBC≤10×109/L and LBC≥35×109/L were designed as critical values to predict fetal lung immaturity and fetal lung maturity,respectively,there was no significant difference in the predictive values between LBC count and L/S ratio(P〉0.05);the predictive values of LBC count was significantly higher than those of foam test(P〈0.01).Conclusion:Amnionic fluid LBC count is superior to routine foam test for prediction of fetal lung maturity;compared with L/S ratio,the method is time-saving and labor-saving,and it is the first choice to predict fetal lung maturity in clinic.