摘要 目的了解新型Innovance D-dimer在Sysmex CS2000(CS2000)全自动凝血分析仪上检测血浆D二聚体(D-dimer,D-D)的性能及其cutoff值的适用性。方法分别使用STA Liatest和Innovance D-D两种试剂在配套机型上检测100例患者的血浆D-D,分析两者的偏差和相关性,并根据排除静脉血栓的cutoff值判断两者符合率。结果在检测范围内两方法具有统计学相关(r=0.9516,P<0.05),不稀释标本(<4 mg/L)检测的相关性也具有统计学意义(r=0.9304,P<0.05)。以0.5 mg/L作为cutoff值,两种方法阴性、阳性和总体符合率分别为92.3%、98.6%和97%。结论Innovance D-D与现行主流方法在临床应用上具有量值可比性和诊断一致性。 Objective To investigate performance of Innovance D-Dimer measuring clinical samples on CS2000 automatic coagulation analyzer with its original cutoff. Method Samples from 100 patients with different diseases were test by STA Liatest D-Direct (with STA Evolution) and Innovance D-dimer respectively,bias between two sets of data and their correlation were analyzed, correspondent rate were calculated according cutoff ruling out venous thromboembolism. Results Data from two methods are correlated statistically not only in analysis range ( r = 0. 9304, P 〈 0.05 ) , but also in undiluted sample with D- dimer lower than 4 mg/L ( r = 0. 9304, P 〈 0.05 ). Negative, positive and whole correspondent rate were 92.3 % ,98.6 % and 97 % taking 0.5 mg/L as cutoff. Conclusion Innovance D-dimer is comparative and correspondent with main stage method in clinical application.
出处 《血栓与止血学》 2012年第3期131-133,共3页 Chinese Journal of Thrombosis and Hemostasis
关键词 D-二聚体 静脉血栓栓塞 血栓形成 D- dimer Venous thromboembolism Thromobosis
  • 相关文献

参考文献12

  • 1DEMPFLE CE. D-dimer: standardization versus harmonization [ J ]. Thromb Haemost ,2006,95 (3) :399-400.
  • 2DEMPFLE CE. D-dimer assays: the current status and newassay technologies [ J ]. Thromb Res,2006,118:569-710.
  • 3BOUNAMEAUX H, SCHNEIDER PA, REBER G,et al. Measurement of plasma D-dimer for diagnosis of deep ve~ous thr,~.~h~sis[ J]o Am J Clin Pathol, 1989,91 : 82-85.
  • 4DEMPFLE CE,ZIPS S, ERGUL H, et al. The Fibrin Assay Compari- son Trial (FACT) : evaluation of 23 quantitative D- Dimer assays as basis for the development of D- Dimer calibrator FACT study group [ J]. Thromb Haemost ,2001,85:671-678.
  • 5GAFFNEY PJ, CREIGHTON LJ, CALLUS, M. et al. Monoclonal anti- bodies to crosslinked fibrin degradation products (XL-FDP). II. Eval- uation in a variety of clinical conditions [ J ]. British Journal of Haem- atology, 1988,68,91 -96.
  • 6ROWBOTHAM BJ, CARROLL P, WHITAKER AN, et al. Measure- ment of crosslinked fibrin derivatives use in the diagnosis of venous thrombosis[ J]. Thromb Haemost, 1987,57:59-61.
  • 7DESIREE CH, MARIJA MRZ. Evaluation of the Innovance D-DIMER analytical performance [ J 1. Clin Chem Lab Med, 2009,47 ( 8 ) : 945- 951.
  • 8REBER G, de MOERLOOSE P. D-Dimer assays for the exclusion of venous thromboembolism [ J ]. Sem Thromb Hemost, 2000, ~6 : 619- 624.
  • 9CHRIS G, CORALIE P, CLAIRE W. An evaluation of rapid D-dimer assays for the exclusion of deep vein thrombosis[ J]. British Journal of Haematology, 128,842 - 848.
  • 10LENNART R. KLAUS D. Comparative evaluation of difterent D-dimer assays for exclusion of venous thromboembolism in a clinical routine setting[ J]. J Lab Med ,2008,32 (3) :215-217.