期刊文献+

Value of a new post-procedural intravascular ultrasound score in predicting target vessel revascularization after coronary drug-eluting stents implantation

Value of a new post-procedural intravascular ultrasound score in predicting target vessel revascularization after coronary drug-eluting stents implantation
原文传递
导出
摘要 Objective:There is no simple or feasible post-procedural intravascular ultrasound (IVUS) score to predict major adverse cardiac events (MACE) in patients undergoing drug-eluting stents (DES) implantation. The aim of this study is to validate a new IVUS score for predicting MACE. Methods:A total of 295 patients (with 322 lesions) were enrolled. IVUS score was calculated in each lesion based on five IVUS morphological characteristics:inflow/outflow disease, malapposition, underexpansion, tissue protrusion, and edge dissection (iMUTE score). We assigned two points to an underexpansion and one point for each presence of other factors. Patients were divided into low score (iMUTE score<2, n=137) and high score (iMUTE score≥2, n=158) groups. Results:At one year follow-up, a trend was seen in favor of the low iMUTE score group in MACE (3.65% vs. 10.10%; P=0.052), and there was more target vessel revascularization (TVR) in the high iMUTE score group compared with low score group (6.96% vs. 1.46%; P=0.044). Low iMUTE score was an independent predictor of freedom from TVR at one year (adjusted hazard ratio (HR) 0.5, 95% confidence interval (CI) 0.1-0.8; P=0.02). Conclusions:Post-procedural IVUS iMUTE scoring was simple and feasible in clinical practice, and can provide independent prognostic value for TVR in patients undergoing DES implantation. Objective: There is no simple or feasible post-procedural intravascular ultrasound (IVUS) score to predict major adverse cardiac events (MACE) in patients undergoing drug-eluting stents (DES) implantation. The aim of this study is to validate a new IVUS score for predicting MACE. Methods: A total of 295 patients (with 322 lesions) were enrolled. IVUS score was calculated in each lesion based on five IVUS morphological characteristics: inflow/outflow disease, malapposition, underexpansion, tissue protrusion, and edge dissection (iMUTE score). We assigned two points to an underexpansion and one point for each presence of other factors. Patients were divided into low score (iMUTE score〈2, n=137) and high score (iMUTE score〉2, n=158) groups. Results: At one year follow-up, a trend was seen in favor of the Low iMUTE score group in MACE (3.65% vs. 10.10%; P=0.052), and there was more target vessel revascularization (TVR) in the high iMUTE score group compared with low score group (6.96% vs. 1.46%; P=0.044). Low iMUTE score was an independent predictor of freedom from TVR at one year (adjusted hazard ratio (HR) 0.5, 95% confidence interval (CI) 0.1-0.8; P=0.02). Conclusions: Post-procedural IVUS iMUTE scoring was simple and feasible in clinical practice, and can provide independent prognostic value for TVR in patients undergoing DES implantation.
出处 《Journal of Zhejiang University-Science B(Biomedicine & Biotechnology)》 SCIE CAS CSCD 2012年第8期624-630,共7页 浙江大学学报(英文版)B辑(生物医学与生物技术)
基金 supported by the Key Project of the National 12th Five-Year Research Program of China (No. 2012ZX09303016-002) the PhD Initiative Research Grant (No. 20091100) of Liaoning Province, China
关键词 Intravascular ultrasound Drug eluting stents Scoring schemes percutaneous coronary intervention Intravascular 超声;药 eluting stents;得分计划;经皮的冠的干预
  • 相关文献

参考文献14

  • 1Cheneau, E., Leborgne, L., Mintz, G.S., Kotani, J., Pichard, A.D., Satler, L.F., Canos, D., Castagna, M., Weissman, N.J., Waksman, R., 2003. Predictors of subacute stent thrombosis: results of a systematic intravascular ultrasound study. Circulation, 108(1):43-47.
  • 2Daemen, J., Wenaweser, P., Tsuchida, K., Abrecht, L., Vaina, S., Morger, C., Kukreja, N., Juni, P., Sianos, G., Hellige, C., et al., 2007. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two- institutional cohort study. Lancet, 369(9562):667-678.
  • 3Fujii, K., Carlier, S.G., Mintz, GtS., Yang, Y.M., Moussa, I., Weisz, G., Dangas, G., Mehran, R., Lansky, A.J, Kreps, E.M., et al., 2005. Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after sirolimus-eluting stent implantation: an intravascular ultrasound study. J. Am. Coll. Cardiol., 45(7):995-998.
  • 4Gerber, R.T., Latib, A., Ielasi, A., Cosgrave, J., Qasim, A., Airoldi, F., Chieffo, A., Montorfano, M., Carlino, M., Michev, I., et al., 2009. Defining a new standard for IVUS optimized drug eluting stent implantation: the PRAVIO study. Catheter. Cardiovasc. Interv., 74(2):348-356.
  • 5Hong, M.K., Mintz, G:S., Lee, C.W., Park, D.W., Choi, B.R.,/ Park, K.H., Kim, Y.H., Cheong, S.S., Song, J.K., Kim, J.J., et al., 2006. Intravascular ultrasound predictors oH angiographic restenosis after sirolimus-eluting stent implantation. Eur Heart J., 27(11):1305-1310.
  • 6Joner, M., Finn, A.V., Farb, A., Mont, E.K., Kolodgie, ED., Ladich, E., Kutys, R., Skorija, K., Gold, H.K., Virmani, R., 2006. Pathology of drug-eluting stentts in humans: delayed healing and late thrombotic risk. d. Am. Coll. CardioL, 48(1):193-202.
  • 7Lee, C.W., Kang,'S.J., Park, D.W., Lee, S.H., Kim, Y.H., Kim] J.J., Park, S.W., Mintz, GS., Park, S.J., 2010{ Intravaseular ultrasound findings in patients with very late stent thrombosis after either drug-eluting or bare-metal stent implantation. J. Am. Coll. Cardiol.I 55(18):1936-1942.
  • 8Morice," M.C., Serruys, EW., Sousa" J.E., Fajadet, J., Ban Hayashi, E., Perin, M., Colombo, A., Schuler, G., Barragan, P., Guagliumi, G~, et al., 2002. A randomized comparison of a sirolimus-eluting stent with a standardstent for coronary revascularization. N. Engl. J. Med., 346(23):1773-1780.
  • 9Moses, J.W., Leon, M.B., Popma, J.J., Fitzgerald, EJ., Holmes, D.R., O'Shaughnessy, C., Caputo, R.P., Kereiakes, D.J., Williams, D.O., Teirstein, ES., et al., 2003. Sirolimus- eluting stents versus standard stents in patients with stenosis in a native coronary artery. N. Engl. J.. Med., 349(14):1315-1323.
  • 10Moses, J.W., Dangas, G., Mehran, R., Mintz, G.S., 2008. Drug-eluting stents in the real world: how intravascular ultrasound can improve clinical outcome. Am. J. Cardiol., 102(9 Suppl.):24J-28J.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部