离子排斥色谱结合光度检测的水质监测系统用于测定离子型营养物(英文) 被引量:2

Water quality monitoring system for determination of ionic nutrients by ion-exclusion chromatography with spectrophotometric detection on cationand anion-exchange resin columns using water eluent
下载PDF
导出
摘要 A unified ion-exclusion chromatography(IEC) system for monitoring anionic and cationic nutrients like NH + 4,NO 2,NO 3,phosphate ion,silicate ion and HCO 3 was developed and applied to several environmental waters.The IEC system consisted of four IEC methodologies,including the IEC with ultraviolet(UV) detection at 210 nm for determining NH + 4 on anion-exchange separation column in OH form connected with anion-exchange UV-conversion column in I form in tandem,the IEC with UV-detection at 210 nm for determining simultaneously NO 2 and NO 3 on cation-exchange separation column in H + form,the IEC with UV-detection at 210 nm for determining HCO 3 on cation-exchange separation column in H + form connected with anion-exchange UV-conversion column in I form in tandem,and the IEC with visible-detection based on molybdenum-blue reaction for determining simultaneously silicate and phosphate ions on cation-exchange separation column in H + form.These IEC systems were combined through three manually-driven 6-port column selection valves to select each separation column to determine selectively the ionic nutrients.Using this sequential water quality monitoring system,the analytical performances such as calibration linearity,reproducibility,detection limit and recovery were also tested under the optimized chromatographic conditions.This novel water quality monitoring system has been applied successfully for the determination of the ionic eutrophication components in sub-urban river waters. A unified ion-exclusion chromatography(IEC) system for monitoring anionic and cationic nutrients like NH + 4,NO 2,NO 3,phosphate ion,silicate ion and HCO 3 was developed and applied to several environmental waters.The IEC system consisted of four IEC methodologies,including the IEC with ultraviolet(UV) detection at 210 nm for determining NH + 4 on anion-exchange separation column in OH form connected with anion-exchange UV-conversion column in I form in tandem,the IEC with UV-detection at 210 nm for determining simultaneously NO 2 and NO 3 on cation-exchange separation column in H + form,the IEC with UV-detection at 210 nm for determining HCO 3 on cation-exchange separation column in H + form connected with anion-exchange UV-conversion column in I form in tandem,and the IEC with visible-detection based on molybdenum-blue reaction for determining simultaneously silicate and phosphate ions on cation-exchange separation column in H + form.These IEC systems were combined through three manually-driven 6-port column selection valves to select each separation column to determine selectively the ionic nutrients.Using this sequential water quality monitoring system,the analytical performances such as calibration linearity,reproducibility,detection limit and recovery were also tested under the optimized chromatographic conditions.This novel water quality monitoring system has been applied successfully for the determination of the ionic eutrophication components in sub-urban river waters.
出处 《色谱》 CAS CSCD 北大核心 2012年第7期721-727,共7页 Chinese Journal of Chromatography
基金 financially supported from TOSOH Corporation and Fellowships Program for Young Scientists of the Japan Society for the Promotion of Science on FY2011
关键词 ion-exclusion chromatography(IEC) anionic and cationic nutrients column selection valve water eluent ion-exclusion chromatography(IEC) anionic and cationic nutrients column selection valve water eluent
  • 相关文献

参考文献14

  • 1JISK0102-2008.
  • 2JISK0101-1998.
  • 3Mori M, Itabasi H, Ikedo M, et al. Talanta, 2006, 70 : 174.
  • 4Tanaka K, Ishizuka T, Sunahara H. J Chromatogr, 1979, 177 : 21.
  • 5Tanaka K. Bunseki Kagaku, 1982, 31 : T106.
  • 6Mori M, Tanaka K, Helaleh M I H, et al. J Chromatogr A, 2003, 997 : 191.
  • 7Tanaka K. Bunseki Kagaku, 1981, 30:661.
  • 8Ikedo M, Mori M, Karachi K, etal. Anal Sci, 2006, 22:117.
  • 9Nakatani N, Kozaki D, Masuda W, et al. Anal Chim Acta, 2008, 519:110.
  • 10Nakatani N, Masuda W, Kozaki D, et al. Anal Sci, 2009, 25 : 379.