国产回弹式眼压计临床应用对比分析 被引量:1

Clinical evaluation of SUOER rebound tonometer
原文传递
导出
摘要 目的探讨国产回弹式眼压计临床应用的可靠性、稳定性及对特殊病例测量的优越性。方法分别使用国产SW-500 SUOER回弹式眼压计、ICARE回弹式眼压计及非接触眼压计(NCT)测量眼科门诊就诊124例患者(124只眼)眼压,并对三者测量结果的相关性进行对比分析。结果SUO-ER、ICARE、NCT三种眼压计测量的均值分别为:(16.82±4.46)mmHg、(16.90±5.50)mmHg、(16.36±5.16)mmHg,SUOER与ICARE测量结果相比较差异无统计学意义(P=0.67),NCT与其他两种眼压计相比差异均有统计学意义。三种眼压计测量结果彼此之间具有良好的相关性分别为0.917(NCT—ICARE)、0.927(NCT—SUOER)、0.944(ICARE—SUOER)。不同眼压区间三种测量值在中眼压组中两两之间的差值较小,在低眼压组与高眼压组中差值较大。不同操作者使用SUOER回弹式眼压计测量结果差异无统计学意义。SUOER回弹式眼压计所需检查时间平均为6.12s,小于NCT平均测量时间10.37秒,77%的被检查者认为其舒适度优于NCT。结论国产SUOER回弹式跟压计与ICARE回弹式眼压计一样能方便准确舒适地应用于各种临床患者,尤其对于测量婴幼儿等不能固视、配合不良及瘫痪不能坐位测量的患者有一定的优越性。 Objective To approach the confidence, stability and superiority of SUOER rebound tonometer. Methods Intraocular pressure (IOP) measurements were performed on 124 cases (124 eyes) with SUOER rebound tonometer, ICARE rebound tonometer and NCT tonometer respectively. Correlation of the results measured by three types of tonometer was analyzed. Results The mean lOP measured by SUOER, ICARE, and NCT were 16.82±4.46mmHg, 16.90±5.50mmHg, 16.36± 5.16mmHg, respectively. No significant difference was found between the result of SUOER and ICARE, and there was a significant difference between the results of NCT and other methods. Significant correlation could be found among all three methods with the correlation coefficient of 0.917 (NCT-ICARE), 0.927(NCT-SUOER), 0.944 (ICARE-SUOER). The differences of lOP among three types of tonometers were smaller in medium lOP group compared with high and low lOP group. No significant inter-observer difference was found by SUOER tonometer. The average examine dura- tion was 6.12 seconds by SUOER tonometer, which was shorter than that of NCT (10.37 seconds). 77% patients felt more comfortable with SOUER tonometer compared with NCT. Conclusion SOUER tonometer show well accuracy, convenience, and compliance for clinical use for various pa- tients, especially for the special patients such as infants, patients without good fixation, bed-ridden patients.
出处 《中国实用眼科杂志》 CSCD 北大核心 2012年第8期980-982,共3页 Chinese Journal of Practical Ophthalmology
关键词 回弹式眼压计 眼压 眼压测量 Rebound tonometer Intraocular pressure Intraocular pressure measurement
  • 相关文献

参考文献10

  • 1张宇燕,孙兴怀.常见眼压计的临床评价[J].眼视光学杂志,2004,6(2):132-135. 被引量:31
  • 2石晶明,蒋幼芹.非接触眼压计测量的评价——与Goldmann眼压计的比较[J].中国实用眼科杂志,2002,20(5):370-372. 被引量:28
  • 3Kontiola AL.A new electromechanical method for measuring in- traocular pressure[J].Doc Ophthalmol, 1997,93(3) :265-276.
  • 4Schreiber W,Vorwerk CK,Langenbucher,et al.A comparison of re- bound tonometry (ICare)with TonoPenXL and Goldmann applana- tion tonometry [ J ] .Ophthalomologe,2007,104(4):299-304.
  • 5刘苏冰,聂晓丽,刘凤云,唐秀侠,辛宝莉.LASIK前后非接触性眼压测定值的分析[J].眼科研究,2001,19(2):176-177. 被引量:4
  • 6Chui WS,Lam A,Chen D,et al.The influence of corneal proper- ties on rebound tonometry[J].Ophthalmolopy,2008,115(1)$0-84.
  • 7甄毅,王涛,王文欣,郑松.国产回弹式眼压计的临床评价[J].眼科新进展,2010,30(6):555-559. 被引量:8
  • 8Davies LN,Bartlett H,Mallen EA,et al.Clinical evaluation of re- bound tonometer [J].Acta: Ophthalmic Physiol Opt, 2005, 25: 436-440.
  • 9Sahin A,Basmak H,Niyaz L,et al.Reproducibility and toberabitl- ity of the ICare rebound tonometer in school children[J]. J Glaucoma, 2007, 16(2) : 185-188.
  • 10Moreno-Montanes J, Garica N,Fernandez-Hortelano A,et al.Re- bound Tonometer Compared With Goldmann Tonometer in Nor- mal and Pathologic Corneas[J].Comea,2007,26(4):427-430.

二级参考文献47

  • 1Pandav SS,Sharma A,Gupta A,et al.Reliability of ProTon and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984.
  • 2Garzoz HJ,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy a comparison of three different tonometers[J].Cornea,2001,20(1):33-36.
  • 3Duch S,Serra A,Castanera J,et al.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma,2001,10(4):261-265.
  • 4Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794.
  • 5Munoz LA,Perez AL,Rahhal MS,et al.Assessment of applanation tonometry after hyperopic laser in situ keratomileusis[J].Cornea,2002,21(2):156-160.
  • 6Copt RP,Tomas R,Mermoud A.Corneal thickness and normal tension glaucoma[J].Arch Ophthalmol,1999,117(1):14-16.
  • 7Dekker PW,Robert YCA,Kanngiesser H,et al.Principles of contact lens tonometry[J].International Ophthalmology,1998,22(2):105-111.
  • 8Shah S,Chatterjee A,Mathai M,et al.Relationship between corneal thickness and messured intraocular pressure in a general opthalmology clinic[J].Ophthalmology,1999,106(11):2154-2160.
  • 9Ventura ACS,Bohnke M,Mojon DS.Central corneal thickness measurements in patients with normal tension glaucoma,primary open angle flaucoma,pseudoexfoliation glaucoma,or ocular hypertension[J].Br J Ophthalmol,2001,85(6):792-795.
  • 10Brandt JD,Beiser JA,Kass MA,et al.Central corneal thickness in the ocular hypertension treatment study(OHTS)[J].Ophthalmology,2001,108(10):1779-1785.

共引文献60

同被引文献9

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部