Lenstar LS 900与角膜超声测厚仪测量角膜中央厚度的对比观察 被引量:3

A comparative investigation of Lenstar LS 900 and A ultrasound pathymetry for measurement of central corneal thickness
下载PDF
导出
摘要 目的比较lenstar LS 900与角膜超声测厚仪测量人中央角膜厚度(CCT)的差异,为临床检查提供参考。方法选取屈光手术者26例(52只眼),分别用lenstar LS 900和角膜超声测厚仪测量双眼CCT,两种仪器均测量3次或以上,取其平均值,对所得数据进行统计学分析。结果 52只眼CCT测量结果:lenstar LS 900为(538.27±40.38)μm,A超为(534.15±41.12)μm。两种测量方法结果差异有统计学意义(t=2.9643,P<0.005)。但测量结果高度相关(r=0.970),前者所测CCT较后者厚(4.11±10.11)μm。结论 lenstar LS 900与角膜超声测厚仪对CCT的测量结果呈较高的相关性,两种仪器具有较高的测量准确性,lenstar LS 900可以作为角膜厚度的测量工具之一。 Objective To compare two method to measure central corneal thickness and to provide a reference for clinical examination. Methods Measure central corneal thickness of 26 patients ( 52 eyes ) before the refractive operation respectively with lenstar LS 900 and A ultrasound pathymetry. Two instnunents were measured for 3 times or more, take the average value, the data obtained were statistically analyzed. Results Central corneal thickness value:LS 900 (538.27 ± 40.38) μm,A ultrasonic (534. 15 ± 41.12) μm. Two methods for measuring result difference has statistics significance ( t = 2. 9643, P 〈 0. 005 ). But the measurement results were highly correlated ( r = 0.970 ). The former is ( 4. 11 ± 10. 11 ) μm thicker than the latter. Conclusion Lenstar LS 900 and A ultrasound pathymetry showed a high correlation. Both of the instruments have high measuring accuracy. Lenstar 900 can be used as a tool of measuring corneal thickness.
出处 《临床眼科杂志》 2012年第4期307-309,共3页 Journal of Clinical Ophthalmology
关键词 角膜超声测厚 中央角膜厚度 LENSTAR LS 900 A ultrasound pathymetry Central corneal thiekness Lenstar LS900
  • 相关文献

参考文献6

  • 1华焱军,黄锦海,王勤美.角膜厚度的临床意义及测量方法进展[J].国际眼科杂志,2011,11(8):1376-1378. 被引量:14
  • 2李坤,陆蓓,崇晓霞,陈彦玲,李晓玲.麻醉剂和泪膜对健康人角膜厚度的影响[J].眼视光学杂志,2004,6(4):244-246. 被引量:16
  • 3黄磊,段宝萍,胡琦.角膜厚度的非接触式测量及研究进展[J].临床眼科杂志,2010,18(4):380-383. 被引量:3
  • 4Tappeiner, C. Rohrer, K. Frueh, et al. Clinical comparison of bi- ometzy using the non-contact optical low coherence reflectometer ( Lenstar LS 900 ) and contact ultrasound biometer ( Tomey AL- 3000) in cataract eyes. BrJ Ophthalmol, 2010, 949:666-4567.
  • 5El Chehab H, Giraud JM, I~ Cone A, et al. Comparison betweenLenstar LS 900 non-contact biometay and OcuScan RXP contact bi- ometry for task delegation . J Fr Ophtalmol,2011,34 : 175-80.
  • 6Rainer G, Pettemel V, Findl O, et al. Comparison of ultrasound paohymetry and partial coherence interferemetry in the measure- ment of central comesl thickness. J Cataract Refract Surg,2002, 28:2142-2145.

二级参考文献44

  • 1吴勤,段宣初,蒋幼芹,卿国平,江冰,石晶明.非接触式角膜内皮镜测量正常角膜厚度的研究[J].眼科学报,2004,20(4):229-232. 被引量:12
  • 2葛坚,刘炳乾,高前应.角膜厚度与青光眼危险性的关系[J].眼科,2005,14(3):145-146. 被引量:29
  • 3黄潇,魏锐利.纳米技术在眼科基础研究中的应用[J].眼科新进展,2006,26(10):794-795. 被引量:3
  • 4曾阳发,刘杏,蔡小于,黄晶晶,郑小萍,李媚,王涛,毛真.AC-OCT与OCT测量中央角膜厚度的比较[J].中山大学学报(医学科学版),2007,28(2):230-233. 被引量:9
  • 5Ho JD,Liou SW,Tsai RJ,et al.Estimation of the effective lens position using a rotating Scheimpflug camera.J Cataract Refract Surg,2008,34:2119.
  • 6Giacomo Savini,Piero Barboni,Vincenzo Profazio,et al.Corneal power measurements with the Pentacam Scheimpflug camera after myopic excimer laser surgery.J Cataract Refract Surg,2008,34:809.
  • 7Kawana K,Tokunaqa T,Miyata K,et al.Comparison of corneal thickness measurements using Orbscan II,non-contact specularmicroscopy,and ultrasonic pachymetry in eyes after laser in situ keratomileusis.Br J Ophthalmol,2004;88:466.
  • 8Hiroki Murase,Akira Sawada,Kiyofumi Mochizuki,et al.Effects of corneal thickness on intraocular pressure measured with three different tonometers.Jpn J Ophthalmol,2009,53:1.
  • 9Cheng AC,Rao SK,Lau S,et al.Central corneal thickness measurements by Ultrasound,Orbscan II,and Visante OCT after LASIK for myopia.J Refrac Surg,2008,24:361.
  • 10Kiddee W,Horattanareung O.Intraoperator Repeatability and interoperator reproducibility in the Ultrasonic pachymetry measurements of central corneal thickness.Med Assoc Thai,2009,92:672.

共引文献30

同被引文献54

  • 1Buckhurst PJ,Wolffsohn JS,Shah S.A new optical low coherence reflectometry device for ocular biometry in cataract patients[J] .Br J Ophthalmol,2009,93(7):949-953.
  • 2Koktekir BE,Gedik S,Bakbak B.Comparison of central corneal thickness measurements with optical low-coherence reflectometry and ultrasound pachymetry and reproducibility of both devices[J] .Cornea,2012,31(11):1278-1281.
  • 3Zhao J,Chen Z,Zhou Z.Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devices in myopes[J] .Clin Exp Optom,2013,96(1):92-99.
  • 4Gursoy H,Sahin A,Basmak H.Lenstar versus ultrasound for ocular biometry in a pediatric population[J] .Optom Vis Sci,2011,88(8):912-919.
  • 5Tappeiner C,Rohrer K,Frueh BE.Clinical comparison of biometry using the non-contact optical low coherence reflectometer (Lenstar LS900) and contact ultrasound biometer (Tomey AL-3000) in cataract eyes[J] .Br J Ophthalmol,2010,94(5):666-667.
  • 6Cruysberg LP,Doors M,Verbakel F.Evaluation of the Lenstar LS 900 non-contact biometer[J] .Br J Ophthalmol,2010,94(1):106-110.
  • 7Jasvinder S,Khang TF,Sarinder KK.Agreement analysis of LENSTAR with other techniques of biometry[J] .Eye (Lond),2011,25(6):717-724.
  • 8Bjelo(s) Ronevic M,Bu(s)ic M,Cima I,et al.Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation[J] .Graefes Arch Clin Exp Ophthalmol,2011,249(1):69-75.
  • 9Mylonas G,Sacu S,Buehl W.Performance of three biometry devices in patients with different grades of age-related cataract[J] .Acta Ophthalmol,2011,89(3):e237-e241.
  • 10Salouti R,Nowroozzadeh MH,Zamani M.Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power calculation[J] .Optometry,2011,82(3):140-147.

引证文献3

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部