摘要
目的:研究正常成人胼胝体磁共振轴位扩散张量成像(DTI)与矢状位DTI各向异性(FA)可能的差异,探讨2种测量方法的可靠性。方法:对68例健康志愿者分别行轴位和矢状位DTI扫描,年龄18~78岁。分别测量胼胝体膝部及压部的FA值。对相同部位的轴位和矢状位测量结果进行比较,并行相关性分析。结果:矢状位上测量的胼胝体膝部FA值大于轴位,且两者之间差异有统计学意义(t=-5.833,P<0.001)。胼胝体压部FA值在矢状位与轴位上差异无统计学意义(t=0.435,P>0.05)。结论:矢状位与轴位DTI测量的胼胝体膝部及压部FA值不同,矢状位DTI能研究胼胝体各个区域各向异性的变化。
Objective: To study the possible difference of anisotropy of corpus callosum in mid-sagittal and transverse diffusion tensor imaging(DTI) in normal adults. Methods: Sixty-eight health adult volunteers were collected continuously, the male and female were 35 and 33 respectively, with an age-range from 18 to 78 years. All cases were performed mid-sagittal and trans- verse DTI separately. The fractional anisotropy (FA) values of the genu and splenium of the corpus callosum were measured in FA map. The FA values were analyzed. Results: The FA values of genu was higher in mid-sagittal DTI than that in transverse DTI,and had statistically difference( t = -5. 833, P 〈0. 001). The FA values of splenium in mid-sagittal DTI and that in transverse DTI was no statistically difference( t =0. 435, P 〉0.05). The FA value of genu was higher than that of splenium in mid-sagittal DTI. The FA value of splenium was higher than that of genu in transverse DTI. Conclusion: The FA values of ge-nu and splenium were different in mid-sagittal and in transverse DTI, the anisotropy of different regions of corpus callosum could be measured in mid-sagittal DTI.