摘要
对日本真空理工株式会社制造的 DL- 1 50 0热膨胀仪的测量结果的系统偏差进行了研究。使用该会社提供的铂、Al2 O3标准样品在 50~ 1 2 0 0°C进行测试 ,通过对测试数据的比较分析 ,发现正确选用计算公式和参照标样可有效地降低测量结果的系统偏差。
The system deviation of the measurement results by using a DL 1500 thermal expansion instrument made by Japanese Vacuum Technique Corporation was analysed in this work. Through the measurement of Pt and Al 2O 3 standard samples provided by this corporation in 50~1200°C and subsequent date analysis, it was found that the system deviation of the measurement results can be effectively decreased by using proper calculation formula and reference samples.
出处
《理化检验(物理分册)》
CAS
2000年第8期352-354,共3页
Physical Testing and Chemical Analysis(Part A:Physical Testing)
关键词
热膨胀仪
平均线膨胀系数
系统偏差
金属
Thermal expansion instrument Average linear expansion coefficient System deviation