气相色谱用微波等离子体炬原子发射光谱检测器的Cl,Br,I响应特性的研究 被引量:1

Studies on Response Characteristics of Cl,Br,I of Microwave Plasma Torch Atomic Emission Detector for Gas Chromatography
下载PDF
导出
摘要 将微波等离子体炬原子发射光谱检测器 ( MPT- AED)用作气相色谱检测器 ,采用氩气作为工作气、载气和尾吹气 ,优化了气体流量参数 ;为了防止被测物质在管壁上的沉积 ,考察了氧气作为气相色谱 -微波等离子体炬 -原子发射光谱检测器 ( GC- MPT- AED)清洗气时对 Cl,Br,I 3种元素发射信号的影响。为了考察检测器的分析性能 ,研究了 GC- MPT- AED对有机化合物中 Cl,Br,I3种元素的检出限、线性范围、精密度及其响应特性 ,并将其结果与气相色谱 -电感耦合等离子体 -原子发射光谱检测器 ( GC- ICP- AED)的结果作了比较 ,结果表明该仪器对 Cl,Br的检测能力优于 GC- ICP- AED的结果 ,线性范围与 GC- ICP- The present paper deals with a microwave plasma torch atomic emission detector for gas chromatography Argon is used as support gas, carrier gas and make up gas. The effect of oxygen scavenger gas on the detection performance for chlorine, bromine and iodine is discussed. Detection limits, dynamic ranges, relative standard deviations and response characteristics of GC MPT AED for chlorine, bromine and iodine in organic compounds were studied. The results are favorable in comparing with GC ICP AED.
机构地区 吉林大学化学系
出处 《色谱》 CAS CSCD 北大核心 2000年第3期237-240,共4页 Chinese Journal of Chromatography
基金 国家自然科学基金重点资助项目! (项目批准号 :2 9735160 )
关键词 气相色谱 响应特性 MPT-AED 氯溴 检测 gas chromatography microwave plasma torch atomic emission detector response characteristic
  • 相关文献

参考文献4

  • 1Jin Q,Spectrochimica Acta.B,1991年,46B期,417页
  • 2Huang Y,J Anal Atom Spectrom,1990年,5卷,115页
  • 3Jin Q,J Anal Atom Spectrom,1990年,5卷,487页
  • 4McCormack A J,分析化学,1965年,37卷,1470页