期刊文献+

The Design and Optimization of a Neutron Time-of-Flight Spectrometer with Double Scintillators for Neutron Diagnostics on EAST 被引量:1

The Design and Optimization of a Neutron Time-of-Flight Spectrometer with Double Scintillators for Neutron Diagnostics on EAST
下载PDF
导出
摘要 Neutron energy spectrometry diagnosis plays an important role in magnetic con- finement fusion. A new neutron time-of-flight (TOF) spectrometer with double scintillators is designed and optimized for the EAST toknmak. A set of optimM parameters is obtained by Monte Carlo simulation, based on the GEANT4 and ROOT codes. The electronic setup of the measurement system is designed. The count rate capability is increased by introducing a flash ADC. The designed spectrometer with high resolution and efficiency is capable of being applied to fusion neutron diagnostics. Applications in mixed-energy and continuous energy neutron fields can also be considered. Neutron energy spectrometry diagnosis plays an important role in magnetic con- finement fusion. A new neutron time-of-flight (TOF) spectrometer with double scintillators is designed and optimized for the EAST toknmak. A set of optimM parameters is obtained by Monte Carlo simulation, based on the GEANT4 and ROOT codes. The electronic setup of the measurement system is designed. The count rate capability is increased by introducing a flash ADC. The designed spectrometer with high resolution and efficiency is capable of being applied to fusion neutron diagnostics. Applications in mixed-energy and continuous energy neutron fields can also be considered.
出处 《Plasma Science and Technology》 SCIE EI CAS CSCD 2012年第7期675-682,共8页 等离子体科学和技术(英文版)
基金 supported by the State Key Development Program for Basic Research of China(Nos.2008CB717803,2009GB107001,and2007CB209903) the Research Fund for the Doctoral Program of Higher Education of China(No.200610011023) the National Natural Science Foundation of China(No.10875002)
关键词 magnetic confinement fusion neutron diagnostics TOF spectrometer GEANT4 flash ADC magnetic confinement fusion, neutron diagnostics, TOF spectrometer GEANT4, flash ADC
  • 相关文献

参考文献43

  • 1Wolle B. 1999, Physics Reports, 312:1.
  • 2Zheng S, Chen M, Li J, et al. 2007, Nucl. Fusion, 47: 1053.
  • 3Fischer U, Pereslavtsev P, Hermsmeyer S. 2005, Fusion Eng. Des., 75-79:751.
  • 4Jarvis O N. 1994, Plasma Phys. Control. Fusion, 36: 209.
  • 5Jarvis O N. 2002, Nucl. Instr. Meth. A, 476:474.
  • 6Kallne J, Gorini G, Ballabio L. 1997, Rev. Sci. In- strum., 68:581.
  • 7Okada K, Kondo K, Sato S, et al. 2006, Rev. Sci. In- strum., 77:10E726.
  • 8Elevant T, Aronsson D, Belle P V, et al. 1991, Nucl. Instr. Meth. A, 306:331.
  • 9Gorini G, Kallne J, Ballabio L. 1997, Rev. Sci. In- strum., 68:561.
  • 10Ciacomelli L, Hjalmarsson A, Sjostrand H, et al. 2005, Nucl. Fusion, 45:1191.

同被引文献6