磁共振射频线圈单元数对图像信噪比的影响 被引量:9

Impact of MR Coils Element Numbers on Image Signal to Noise Ratio
下载PDF
导出
摘要 在磁共振成像过程中,射频线圈用来接收信号,对成像的质量起着至关重要的作用。用户在采购磁共振设备时往往用线圈的通道数和单元数来评价一个线圈的成像能力,而线圈的通道数又必须和系统的接收通道数相匹配,因此一般认为,在系统其他条件一定的情况下,线圈的单元数越多,成像质量越好。本文以Philips Achieva3.0T磁共振为平台,分别选取了8通道8单元的Sense Head(头部)线圈和16通道18单元的Sense Neurovascular(神经血管)线圈作为比较对象进行扫描,比较两者成像的信噪比。另外,选取15通道15单元Sense Spine(全脊柱)线圈,先后用不同数量的采集单元数成像,然后对信噪比进行评价。 Radio-frequency (RF) coils are used to receive signals from human bodies during MRI imaging, they are important to the image quality. In the process of purchasing, technicians usually evaluate the imaging quality of RF coils based on their channels and elements numbers. Since the number of coil channels have to be matched with that of MRI system, more elements of RF coils result in better imaging quality under the same condition. In this paper, we chose Philips Achieva3.0T as the platform and compared the signal noise ratios of Sense Head Coil with 8 channels and 8 elements Sense NV coil with 16channels and 18elements. Moreover, we selected 15 channels Sense Spine coil but different number of elements respectively to scan phantoms and evaluate SNR.
出处 《中国医疗设备》 2012年第8期116-119,共4页 China Medical Devices
关键词 磁共振成像仪 射频线圈 信噪比 头线圈 神经血管线圈 全脊柱线圈 MRI radio-frequency coil signal noise ratio sense head coil sense neurovascular coil sense spine coil
  • 相关文献

参考文献6

二级参考文献31

  • 1Ugurbil K,Garwood M,Ellermann J,et al.Imaging at high magnetic fields:initial experiments at 4T[J].Magn Resort Q,1993,9:259-277.
  • 2Lu H,Nagae-Poetscher LM,Golae X,et al.Routine clinical brain MRI sequences for use at 3.0 Tesla[J].J Magn Reson Imaging,2005,22:13-22.
  • 3Alecci M,Collins CM,Smith MB,et al.Radio frequency magnetic field mapping ofa 3 Tesh birdcage coil:experiment and theoretical dependence on sample properties[J].Magn Reson Med,2001,46:379-385.
  • 4Ibrahim TS,Lee R,Baertlein BA,et al.Blfield homogeneity and SAg.calculations for the birdcage coil[J].Phys Med Biol,2001,46:409-619.
  • 5Bernd Bittersohl,Tomas Huang,Schneider E,et al.High-Resolution MRI of the Triangular Fibro cartilage Complex(TFCC) at 3T:Comparison of Surface Coil and Volume Coil[J].J Magn Reson Imaging,2007,26:701-707.
  • 6Kumar A,Bottomley PA.Optimized quadrature surface coil designs[J].Magn Reson Mater Phy,2008,21:41-52.
  • 7Bin Li,WangJie.The Development and Evaluation of New Neck Coil for GE MR System[A].Conl Pro LEEE Eng Med Biol Soc[C].Shanghai:IEEE--EMBS,2005.6033-6036.
  • 8Kaufman L,Krarner DM,Crooks LE,et al.Measuring Signal-to-Noise Ratios in MR.Imaging[J].Radiology,1989,173:265-267.
  • 9Price RR,Axel L,Morgan T,et al.Quality assurance methods and phantoms for magnetic resonance imaging:report of AAPM nuclear magnetic resonance task group no 1[J].Med Phys,1990,17:287-95.
  • 10Firbank MJ,Coulthard A,Harrison RM,et al.A comparison of two methods for measuring the signal to noise ratio on MR images[J].Phys Med Biol,1999,44(12):N261-264.

共引文献16

同被引文献78

引证文献9

二级引证文献73

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部