PEEK棒椎弓根螺钉内固定系统的三维有限元分析 被引量:5

Finite element analysis on PEEK rods pedicle screw system
下载PDF
导出
摘要 目的通过三维有限元方法,比较PEEK棒与钛棒两种固定系统的椎间应力、螺钉应力、相邻节段椎间盘的生物力学变化。方法建立L3~L5的正常腰椎模型,在Pro/E软件中建立两种椎弓根螺钉系统模型。将两者组合构建L4/5节段内固定系统模型,在同样的约束和载荷条件下,比较屈伸、侧屈、旋转三种运动状态下,植骨融合前后,两种内固定装置的生物力学特点。结果在植入PEEK棒和传统钛棒后,在植骨融合前后两种状态下,头侧L3~L4相邻节段的纤维环和髓核应力PEEK棒组均较钛棒组降低。在植骨融合前PEEK棒组的椎间应力要更接近于正常的生理应力,PEEK棒组内固定螺钉表面的应力在植骨融合前后均较小。结论 PEEK棒与传统钛棒相比具有能减少内固定系统载荷与增加椎间植骨应力刺激、降低骨螺钉界面应力、减轻相邻节段载荷。 Objective To analyze the biomechanical characteristics of PEEK rods and Titanium rods, by comparing the stresses on the inter-vertebrae of the fixation segments and adjacent segments and the surface ofpediele screws using finite element analysis. Method Based on the established finite element model of L3-L~, the biomechanical characteristics of the both systems were compared at flexion-extension, lateral bending and axial rotation before and after interbody fusion with the same constraint and conditions. Results Higher stesses were observed on the inter-vertebrae of the fixation seg- ments and lower stresses were recorded on the inter-vertebrae of the adjacent segments for PEEK rods system under flex- ion-extension, lateral bending and axial rotation. Lower stresses on the pedicle screw were also found for PEEK rods. Conclusion In contrast to the Titanium rods, PEEK rods can support the fusion process by increasing loads on the anterior column bone graft while reducing loads on the bone screw interface, and may also reduce the stesses on the adjacent seg- ments.
机构地区 济南军区总医院
出处 《生物骨科材料与临床研究》 CAS 2012年第4期4-8,65,66,共7页 Orthopaedic Biomechanics Materials and Clinical Study
关键词 PEEK棒 内固定 生物力学 有限元分析 PEEK rods Internal fixation Biomechanical Finite element analysis
  • 相关文献

参考文献20

  • 1Rohlmann A,et al.Effect of an internal fixator and a bone graft onintersegmental spinal motion and intradiscal pressure in the adjac-ent regions.Eur Spine J,2001,301-308.
  • 2Zander T,et al.Estimation of muscle forces in the lumbar spineduring upper-body inclination.2001,73-80.
  • 3Schultz A B,et al.Mechanical poperties of human lumbar spinemotion segments-1.responses in flexion,extension,lateral ben-ding,and torsion.Journal of Biomechanical Engineering,1979,101﹙1﹚:46-52.
  • 4Dooris AP,et al.Load-sharing between anterior and posterior ele-ments in a lumbar motion segment implanted with an artificialdisc.Spine,2001,26﹙6﹚.
  • 5An HS,Singh K,Vaccaro AR,et al.Biomechanical evaluation ofcontemporary posterior spinal internal fixation configurations inan unstable burst-fracture calf spine model:special references ofhook configurations and pedicle screws[J].Spine﹙Phila Pa 1976﹚,2004,29﹙3﹚:257-262.
  • 6Vaccaro AR,Garfin SR.Internal fixation﹙pedicle screw fixation﹚for fusions of the lumbar spine[J].Spine﹙Phila Pa 1976﹚,1995:157-165.
  • 7Park P,Garton HJ,Gala VC,et al.Adjacent segment disease afterlumbar or lumbosacral fusion:review of the literature[J].Spine﹙Phila Pa 1976﹚,2004,29﹙17﹚:1938-1944.
  • 8Hilibrand AS,Robbins M.Adjacent segment degeneration and ad-jacent segment disease:the consequences of spinal fusion?[J]Spi-ne J,2004,4﹙6 Suppl﹚:190-194.
  • 9Goto K,Tajima N,Chosa E.Effects of lumbar spinal fusion on theother lumbar intervertebral level﹙s 3Dfinite element analysis﹚[J].J Orthop Sci,2003,8﹙4﹚:577-584.
  • 10Kurtz SM,Devine JN.PEEK biomaterials in trauma,orthopedic,and spinal implants.Biomaterials[J].2007,28﹙32﹚:4845-4869.

同被引文献29

  • 1郑燕平,刘新宇,贾龙,王延国,黎君彦.腰椎后路椎间融合术后椎间融合的X线片及三维CT评价[J].中华骨科杂志,2009,29(12):1104-1108. 被引量:22
  • 2Kanayama M,Togawa D,Hashimoto T,et al.Motion-preserving surgery can prevent early break down of adjacent segments:comparison of posterior dynamic stabilization with spinal fusion[J].Spinal Disord Tech,2009,22(7):463-467.
  • 3Park P,Garton HJ,Gala VC,et al.Adjacent segment disease after lumbar or lumbosacral fusion:revie of the literature.Spine,2004,29(17):1938-1944.
  • 4Brantigan JW,Steffee AD,Lewis ML,et al.Lumbar interbody fusion using the Brantigan I/F cage for posterior lumbar interbody fusion and the variable pedicle screw placement system:two-year results from a Food and Drug Administration investigational device exemption clinical trial[J].Spine,2000,25(11):1437-1446.
  • 5Kurtz SM,Devine JN.PEEK biomaterials in trauma,orthopedic,and spinal implants[J].Biomaterials,2007,28(32):4845-4869.
  • 6Ponnappan RK,Serhan H,Zarda B,et al.Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation[J].Spine,2009,9(3):263-267.
  • 7Mavrogenis AF,Vottis C,Triantafyllopoulos G,et al.PEEK rod systems for the spine[J].Eur J Orthop Surg Traumatol.2014,24(Suppl 1):S111-116.
  • 8Aota Y,Kumano K,Hirabayashi S.Postfusion instability at the adjacent segments after rigid pedicle screw fixation for degenerative lumbar spinal disorders[J].J Spinal Disord,1995,8(6):464-473.
  • 9Etebar S,Cahill DW.Risk factors for adjacent-segment failure following lumbar fixation with rigid instrumentation for degenerative instability[J].Neurosurg,1999,90(2 Suppl):163-169.
  • 10Athanasakopoulos M,Mavrogenis AF,Triantafyllopoulos G,et al.Posterior spinal fusion using pedicle screws[J].Orthopedics,2013,36(7):e951-e957.

引证文献5

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部