摘要 目的通过对不同年份介入治疗过程中不同患者所受辐射剂量的分析,探讨影响辐射剂量的主要因素。方法应用热释光测量法,对不同疾病患者体表受照剂量进行测量、统计分析。结果介入操作中,患者受照剂量差别很大,受多种因素(设备性能、病种、手术复杂程度、术者的操作技术等)影响。平均体表剂量中肝癌患者后背正中最高,为(142.92±75.56)mGy,其次为肿瘤患者。个体受照剂量最高为1 097.00 mGy,为骶尾部脊索瘤患者。2011年患者受照剂量较2009年呈不同程度的上升。结论应采取各种有效措施使患者的受照剂量合理降低,特别是主动防护。 [Objective]To investigate the radiation dose of patients during therapeutic interventional procedures,and analyze the main factors which influence radiation dosage.[Methods]Theromoluminescence was used to measure and analyze the X-ray radiation dose at different positions of patients in therapeutic interventional procedures.[Results] The difference of radiation dose were significant during therapeutic interventional procedures due to a variety of factors(equipment performance,type of disease,surgical complexity,the surgeon,operating techniques,etc).The highest average dose of skin in interventional procedures was the middle of back in patients with hepatocellular carcinoma(142.92±75.56) mGy,follwed by tumor patients.The highest individual dose of skin was 1 097.00 mGy.It is the chordoma in sacrococcygeal region.The radiation dose of patients in 2011 was higher than that in 2009.[Conclusion]It is necessary to reduce radiation dose by adopting effective measures,especially active protection..
出处 《职业与健康》 CAS 2012年第14期1728-1729,共2页 Occupation and Health
关键词 介入放射学 患者 不同年度 辐射剂量 辐射防护 Interventional radiology Patients Different years Radiation dosage Radiation protection
  • 相关文献

参考文献6

二级参考文献34

共引文献13

同被引文献44

  • 1樊红霞,张挽时.介入放射与放射性白内障研究进展[J].中华放射医学与防护杂志,2006,26(2):200-202. 被引量:9
  • 2GBZ 106-2002,放射性皮肤疾病诊断标准[S].
  • 3余宁乐,杨小勇,张乙眉,王进,陈群.江苏省介入放射学应用现状研究[J].中国辐射卫生,2007,16(4):450-452. 被引量:4
  • 4UNSCERA, Sources and effects of ionizing radiation [R]. New York: United Nations, 1993.69-71.
  • 5GBZ95-2014.职业性放射性白内障的诊断[s].
  • 6ICRP. Avoidance of radiation injuries from medical interventional pro- cedures[R]. Ann ICRP: 2000.30(2):7-67.
  • 7Miller DL, Baiter S, Cole PE. Radiation dose in interventional radiolo- gy procedures: the RAD-IR study: part 1: overall measures of dose [J]. J Vasc lnterv Radiol, 2003, 14(6): 711-727.
  • 8Mooney RB, Mekinstry CS, Kamel HA. Absorbed dose and determin- istic effects to patients from interventional neuroradiology[J]. Br J Ra- dio/, 2000, 73(871): 745-751.
  • 9Vano E, Gonzalez L, Beneytez F, et al. Lens injuries induced by oc- cupational exposure in non-optimized interventional radiology laborato- ries[J]. Br J Radiol, 1998, 71(847): 728-733.
  • 10Vano E, Arranz L, Sastre JM, et al. Dosimetric and radiation protec- tion considerations based on some cases of patient skin injuries in in- terventional cardiology[ J ]. Br J Radiol, 1998, 71 ( 845 ) : 510-516.

引证文献3

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部