期刊文献+

伽玛刀立体定向放射外科靶点位置精确度的分析 被引量:2

The analysis of precision and accuracy of target Position in gamma knife stereotactic radiosurgery
下载PDF
导出
摘要 目的 :本文通过测量和分析伽玛刀立体定向放射外科过程中靶点位置误差 ,提出保证伽玛刀治疗质量的建议。材料和方法 :通过胶片法分别测量靶点位置在 Z轴平面和 X轴平面上的总精确度。结果 :靶点在 Z轴和 X轴平面上距离的总精确度分别为 (0 .9± 0 .3) mm和 (0 .8± 0 .2 ) m m。结论 :靶点位置总误差主要来源于定位误差和治疗机的机械误差。定期对 MRI的空间几何失真测试以及结合 CT与 MRI进行靶点立体定位相当必要。 Objective: The paper puts forward the proposals for ensuring quality of the Leksell Gamma Knife(LGK) stereotactic radiosurgery (SRS) by measuring and analyzing the errors of LGK in singular process. [WT5”HZ]Materials and Methods: [WT5”BZ]The precision and accuracy of target position in Z axis plane and X axis plane were measured by the film dosimeter. [WT5”HZ]Results:[WT5”BZ]The tolal precision and accuracy of target position in Z axis plane and X axis plane were(0.9±0.3) mm and (0.8±0.2) mm. [WT5”HZ]Conclusion: [WT5”BZ]The tolal error of target position stems from the spatial distortion of MRI and the graphical coordinate calculation and the mechanical error of LGK.
出处 《中国医学物理学杂志》 CSCD 2000年第2期65-66,共2页 Chinese Journal of Medical Physics
关键词 伽玛刀 立体定向放射外科 定位摆位 靶点 gamma knife stereotactic radiosurgery localization and positioning target precision and accuracy film dosim
  • 相关文献

参考文献7

二级参考文献5

  • 1彭振军.医用磁共振成像技术[M].武汉:湖北科学技术出版社,1997..
  • 2杨开军,颅脑疾病X刀治疗学,1997年,149页
  • 3彭振军,医用磁共振成像技术,1997年,130页
  • 4王肖兰,中华放射肿瘤学杂志,1996年,5期,97页
  • 5王肖兰,张之彬,周汝信,张欣,李培松.X(γ)刀剂量场的测量[J].中华放射肿瘤学杂志,1996,5(2):27-31. 被引量:4