ElektaPrecise直线加速器新型全碳素纤维治疗床床板对放疗剂量的影响 被引量:13

Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose
原文传递
导出
摘要 目的探讨ElektaPrecise直线加速器新型全碳素纤维治疗床床板对放疗剂量的影响。方法应用等中心技术(SAD)测量,设2组射野,均为等中心对穿野(-组穿过床板,另-组不穿过床板),用PTw剂量仪0.6cm3。Farmer电离室比对测量,计算出治疗床主床板、体部延长板、头颈肩延长板不同位置及其衔接处对放疗剂量的衰减。结果6MV能量时,治疗床主床板对剂量的衰减为1.4%~7.2%,主床板衔接头1、4、7和8cm处对剂量的衰减分别为2.8%-38.7%、1.4%-30.1%、1.5%-20.8%和1.4%-11.2%;体部延长板对剂量的衰减为0.5%~5.0%,体部延长板衔接头1cm位置对剂量的衰减为4.7%~15.4%;头颈肩延长板颈部位置对剂量的衰减在0.5%-3.3%,头颈肩延长板肩部位置对剂量的衰减在5.3%-16.7%;延长板和主床板衔接处对剂量的衰减在6.8%~30.4%。结论新型全碳素纤维治疗床床板不同部位对剂量衰减不同,床板衔接部位标记区域对剂量衰减比较大。 Objective To investigate the dosimetric influence of pure carbon fiber treatment tabletop of Elekta Precise new linear accelerator in radiotherapy. Methods Surface-axis distance (SAD) technology was employed for the measurement. Two groups of fields were set and both of them were SAD opposed portals (one of them went through the tabletop, while the other did not). A PTW electrometer and a O. 6 cm3 Farmer ionization chamber were utilized for comparison measurement. Then dose attenuation of the main table board, extended body board, the extended board for head, neck and shoulders, and the joints of these boards were calculated. Results Under the energy of 6 MV, the dose attenuations of the following locations were: 1.4% - 7.2% at the main treatment table board; 2. 8% - 38.7% , 1.4% - 30.1% , 1.5 % -20.8% and 1.4% - 11.2% , respectively at distances of 1,4, ? and 8 cm from the joint of the main table board ;0. 5% - 5.0% at the extended body board; 4. 7% - 15.4% at distance of lcm from the joint of the extended body board; 0. 5% - 3.3% at the neck position of the extended board for head, neck and shoulders; 5.3% - 16.7% at the shoulder positions; and 6.8% - 30. 4% at the joint between the extended boards and the main table board. Conclusions The dose attenuations of the new linear accelerator pure carbon fiber treatment tabletop vary at different locations. Considerable higher attenuations are observed at the table board joints than other locations.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2012年第4期386-389,共4页 Chinese Journal of Radiological Medicine and Protection
基金 贵州省科学技术基金[黔科合(200)号2059]
关键词 治疗床床板 等中心技术 剂量衰减 衰减因子 放疗剂量 Treatment tabletop Source-axis distance technology Dose attenuation Attenuation factor Radiotherapy dose
  • 相关文献

参考文献6

二级参考文献12

共引文献41

同被引文献65

  • 1张旭初,贺秋冬,杨立,肖爱农,康红兵.真空袋对肿瘤吸收剂量的影响[J].湖南师范大学学报(医学版),2006,3(3):54-56. 被引量:3
  • 2盛延兴,隋振忠,郑芳霞,李学章,魏绪国,于书增.立体定向放疗床、头架、真空袋对吸收剂量的影响[J].现代肿瘤医学,2004,12(3):184-184. 被引量:16
  • 3丛也彤,陈大伟,白岚,周银行,朴永峰,王溪,曲雅勤.乳腺癌术后胸壁电子线放射治疗的剂量分布[J].吉林大学学报(医学版),2006,32(5):832-834. 被引量:4
  • 4Spezi E, Ferri A. Dosimetrics of the Siemens IGRT carbon fiber tabletop[J]. Med Dosim, 2007, 32(4): 295-298.
  • 5International Commission on Radiation Units and Measurements. Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures[R]. ICRU Report 24. Washington (DC), 1976: 5.
  • 6Oldham M, L~tourneau D, Watt L, et al. Cone-beam-CT guided ra- diation therapy: A model for on-line application[J]. Radiother Oncol, 2005, 75(3): 271-278.
  • 7Ploquin N, Rangel A, Dunscombe E Phantom evaluation of a commercially available three modality image guided radiation therapy system[J]. Med Phys, 2008, 35(11): 5303-5311.
  • 8Pulliam KB, Howell RM, Followill D, et al. The clinical impact of the couch top and rails on IMRT and arc therapy[J]. Phys Med Biol, 2006,56(23): 7435-7447.
  • 9Njeh CF, Parker J, Spurgin J, et al. A validation of carbon fiber imaging conch top modeling in two radiation therapy treatment planning systems: pbilips pinnacle3 and BrainLAB iPlan RT dose [J]. Radiat Oncol, 2012,7: 190-201.
  • 10Elekta C. Radiation oncology accessories catalog[M]. Stockholm: Elekta company, 2013 : 27 28.

引证文献13

二级引证文献46

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部