期刊文献+

远端联动叶片自锁髓内钉治疗股骨骨折的生物力学研究 被引量:4

The Biomechanical Evaluation of Distal Linkage Vane Intramedullary Self-locking Nail for Femur Fractures
下载PDF
导出
摘要 设计了一种可以远端自锁的髓内钉。对股骨干骨折行远端联动叶片髓内钉固定和交锁钉固定后,对其生物力学进行测试。了解髓内钉治疗股骨骨折的生物力学性能。选6个青壮年男性尸体取双侧完整股骨,中间做成横断骨折,然后用远端联动叶片自锁髓内钉和国产传统交锁髓内钉固定。对股骨-髓内钉复合体进行力学测试。实验结果用SPSS17.0统计软件进行分析。在三点弯曲及扭转实验方面,远端联动叶片自锁髓内钉较国产交锁髓内钉无明显差异。抗轴向载荷时较国产交锁髓内钉差,但满足在肢体非负重状况下生物固定。说明股骨型远端联动叶片髓内钉力学稳定性好,在治疗股骨简单骨折有一定优势。 A remote linkage leaves self-locking intramedullary nail is developed. To test the biomechanical properties of remote linkage intramedullary nail fixation and intramedullary nail fixation after femoral shaft fractures blade, the biomechanical property of intramedullary nailing for femoral fractures is understanded. Six young male body to take the bilateral complete femur, the transverse fracture were made in the middle, and then use the remote linkage leaves self-locking intramedullary nail and domestic conventional interlocking intramedullary fixing. To made a mechanical test of the femur-intramedullary complex, the experimental results with SPSS17.0 statistical software are analyzed. There is no significant differences between remote linkage leaves self-locking intramedullary nail and domestic interlocking intramedullary nail at the three-point bending and torsion experiments, poor resist- ance to axial load than domestic interlocking intramedullary nail, but meet the physically non-weight-bearing status under the biological fixation. It is conclused that femur-type remote linkage blade intramedullary nail has a better mechanical stability, has certain advantages in the treatment of femur fractures.
出处 《科学技术与工程》 北大核心 2012年第26期6585-6590,共6页 Science Technology and Engineering
关键词 股骨骨折 骨折固定术 髓内钉 生物力学 femoral fractures fracture fixation intramedullary nailing biome chanics
  • 相关文献

参考文献6

二级参考文献78

共引文献278

同被引文献29

引证文献4

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部