摘要 对火焰原子吸收光谱法测定运动食品中锌含量进行不确定度评定。通过对该方法人员、设施和环境条件、检测方法及方法确认、设备、测量的溯源性、抽样等方面的全面评估,测量值为36.36mg/kg,其扩展不确定度为1.38mg/kg,K=2。该方法的不确定度较小,方法准确度较高,可用于科学指导功能食品的使用。 The content of iron in functional food was measured by the atomic absorption spectrometry , and the uncertainty was obtained by evaluating the influence of personnel , equipments , environment conditions , testing method , measurement traceability and sampling . The content of iron in functional food was 36 . 3 6mg / kg ,and the expande d uncertainty was 1 . 38mg / kg ,K = 2 . We find the uncertainty is small , and the accuracy is high , which can give scientific guidance for the use of functional food .
出处 《甘肃医药》 2012年第8期564-567,共4页 Gansu Medical Journal
关键词 火焰原子吸收光谱法 功能食品 铁含量 不确定度 AAS functional food iron content uncertainty
  • 相关文献

参考文献4

  • 1中国预防医学科学院营养与食品卫生研究所.食品营养成分测定法[M].北京:人民卫生出版社,1991.169-178.
  • 2中华人民共和国国家标准食品中铁、镁、锰的测定[S]GB/T5009.90-2003.
  • 3中国实验室认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社出版,2002.78.
  • 4中华人民共和国国家计量技术规范.测量不确定度评定与表示[S].JJF.1059-1999.