无源效率刻度法测量高浓铀浓缩度

The Measurement of Enrichment of HEU by ISOCS
下载PDF
导出
摘要 利用HPGe谱仪测量了高浓铀样品γ谱,分析γ谱中235U的特征γ射线163和186 keV以及238U的特征γ射线1 001 keV的峰面积。利用ISOCS软件建立测量模型,对这几条γ射线进行了无源效率刻度。实验结果显示,测量值与参考值相差+1.1%,其误差主要来源于无源效率刻度计算和峰面积计算。 The HEU sample was measured by HPGe,and the spectrum of witch was analysed.The model of HEU and HPGe gamma-ray detector was established.The results shown that the enrichment of HEU sample was +1.1% big than the reference value.The reason was reckoned without the other isotopes in the sample.On the other hand,the simulation of efficient and the calculate of peak area could also cause the error.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2012年第6期650-652,685,共4页 Nuclear Electronics & Detection Technology
关键词 高浓铀 浓缩度 无源效率刻度 HEU enrichment ISOCS
  • 相关文献

参考文献4

  • 1Cong Tam Nguyen, Jozsef Zsigrai. Basic characterization of :tighly enriched uranium by gamma spectrometry[J]. Nuclear Instruments and Methods in Physics Research B,2006,246(2) : 417 -424.
  • 2H. Y eel, M. A. etiner, H. Demirel. Use of the 1001 keV peak of ^234Pa daughter of ^238U in measurement of uranium concentration by HPGe gamma - ray spectrometry [ J ]. Nuclear Instruments and Methods in Physics Research A, 1998,413 ( 1 ) :74 -82.
  • 3H. L. Scott, K.W. Marlow. Gamma - ray emission probabilities of the daughters of ^238U [J]. Nucl. Instr. and Meth. A,1990, 286(3) :549 -555.
  • 4Canberra HPGe manual [ Z ]. Detector specifications and performance data,2001.