期刊文献+

基于RS-485总线的双G-M计数管γ剂量率探测器研制 被引量:4

Development of γ Dose-rate Detector with Two G-M Counter Tubes Based on RS-485 Bus
下载PDF
导出
摘要 根据放射性场所工作人员辐射防护需求,设计了γ剂量率探测器。介绍了探测器硬件电路及软件设计,利用双G-M计数管覆盖高低量程范围,采用RS-485通信接口,可连接最多128个节点,具有远程传输能力,适用于多点分布式实时监测。探测器结构紧凑,功耗低,抗干扰能力强,工作稳定可靠。 The γ dose-rate detector is designed to meet the requirement of radiation protection for people who work in radioactive environment.Two G-M counter tubes are used to cover wide range.The detector has RS-485 communication interface and the signal can be transmitted with long distance.The maximum number of detectors connected with the system is 128.The detector can be used for distributed and real time multi-site monitoring.The detector has compact structure,low power-consume,better anti-interference ability and high reliability.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2012年第6期736-740,共5页 Nuclear Electronics & Detection Technology
关键词 RS-485总线 G-M计数管 Γ剂量率 探测器 RS-485 bus G-M counter tube γ dose-rate detector
  • 相关文献

参考文献7

二级参考文献21

  • 1刘伯学,毛用泽,马文帮,张俊生,李湘葆,叶志跃.G-M计数管剂量率仪用于环境辐射水平测量时的一些问题[J].核电子学与探测技术,1995,15(2):109-113. 被引量:4
  • 2岳清宇 等.低大气层中宇宙射线电离量的分布测量[J].辐射防护,1988,8(6):401-401.
  • 3杨钦元,李瑞香等.建筑材料产品及建材用工业废渣放射性物质控制要求GB6763-2000[S].北京:中国标准出版社,2000.
  • 4张明等.天然石材产品放射防护分类控制标准JC518-93[s].国家建筑材工业局,1993.
  • 5HJ/T61-2001.中华人民共和国环境保护行业标准:辐射环境监测技术规范.[S].,..
  • 6Saez-Vergana J C. Recent Developments of passive and Active Detectors used in the Monitorirg of External Environmental Radiation. Radiat. Prot. Dosim.,2000.92: 83
  • 7Thompson I M G, Andersen C E, Btter-Jensen L ,et al. An International Intercomparison of National Network Systems used to provide Early Warning of a Nuclear Accident Having Transboundary Implications. Radiat.Prot.Dosim.,2000.92:89
  • 8杨本.GM-B型环境γ连续监测仪研制(内部资料)[Z].,1985..
  • 9于群.原子核物理实验方法[M].北京:人民教育出版社,1961.75.
  • 10JJG521-88.中华人民共和国国家计量检定规程:环境监测用X、γ辐射空气吸收剂量率仪.[S].,..