回弹式眼压计在先天性青光眼患者眼压测量中的临床评价 被引量:2

Clinical evaluation of the rebounder tonometer in patients with congenital glaucoma
原文传递
导出
摘要 目的通过与手持压平式眼压计和Schi觟tz眼压计进行比较评价回弹式眼压计在先天性青光眼患者眼压测量中的价值。设计诊断技术评价。研究对象北京同仁眼科中心连续接受眼压测量的平均年龄(12.0±3.0)个月的先天性青光眼患者29例(29眼)。方法应用回弹式眼压计、Perkins手持压平式眼压计、Schi觟tz眼压计测量眼压,每种眼压计测量眼压重复三次。主要指标眼压值,Pearson相关系数(r)。结果回弹式眼压计、Schi觟tz眼压计和压平式眼压计所测眼压平均值分别为(19.3±8.5)mm Hg、(21.1±10.8)mm Hg和(18.5±7.8)mm Hg(F=0.61,P=0.55)。回弹式眼压计与手持压平眼压计的平均差值为(0.8±2.4)mm Hg。Schi觟tz眼压计与手持压平眼压计平均差值为(2.6±2.4)mm Hg,Schi觟tz眼压计与回弹眼压计的平均差值为(1.8±2.4)mm Hg。回弹式眼压计与手持压平式眼压计(r=0.934,P=0.000)、Schi觟tz眼压计(r=0.933,P=0.000)眼压值均显著相关。结论回弹式眼压计的准确性可满足先天性青光眼患者眼压测量的需求。 Objective To evaluate the clinical value of the rebounder tonometer (RBT) in congenital glaucoma by comparing with Perkins applanation tonometer and Schititz tomometer. Design Evaluation of diagnostic techniques. Participants 29 cases (29 eyes) of the congenital glaucoma patients aged 12.0 ± 3.0 months who were measured intraoeular pressure (lOP) in Beijing Tongren Eye Center. Method lOP was measured with the rebound tonometer, Schititz tonometer and Perkins applanation tonometer. Measurements were re- peated three times each tonometer. Main Outcome Measures lOP values and Pearson correlation coefficient. Results Mean lOP values obtained by RBT, Schititz and Perkins tonometer were 19.3±8.5 mm Hg, 21.1±10.8 mm Hg, and 18.5±7.8 mm Hg, respectively (F=0.61, P=0.55). The difference of IOP values among three tonometers was not significant. The average difference between RBT and Perkins tonometer was 0.8±2.4 mm Hg. The average difference between Schioetz and Perkins tonometer was 2.6±2.4 mm Hg. The average differ- ence between Schititz and RBT was (1.8+2.4)mm Hg. The values of Perkins applanation tonometer were significant correlated with RBT (r=0.934,P=-0.000) and Sehititz (r=0.933,P=-0.000). Conclusion The accuracy of rebounder tonometer can meet the IOP measurement needs in congenital glaucoma patients. (Ophthalmol CHN, 2012, 21: 278-281)
出处 《眼科》 CAS 2012年第4期278-281,共4页 Ophthalmology in China
关键词 眼压 先天性青光眼 回弹式眼压计 准确性 intraocular pressure congenital glaucoma rebounder tonometer accuracy
  • 相关文献

参考文献14

  • 1Aponte EP, Diehl N, Mohney BG. Incidence and clinical charac- teristics of childhood glaucoma: a population-based study. Arch Ophthalmol, 2010, 128: 478-482.
  • 2Beck AD. Primary congenital glaucoma in the developing world. Ophthalmology, 2011, 118: 229-230.
  • 3Schneider E, Grehn F. Intraocular pressure measurement-compari- son of dynamic contour tonometry and goldmann applanation tonometry. J Glaucoma, 2006, 15: 2-6.
  • 4Schreiber W, Vorwerk CK, Langenbucher A, et al. A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann appla- nation tonometry. Opbthalmologe, 2007, 104: 299-304.
  • 5Asrani S, Chatterjee A, Wallace DK, et al. Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring de- vice. J Glaucoma, 2011, 20: 74-79.
  • 6Lundvall A, Svedberg H, Chen E. Application of the ICare rebound tonometer in healthy infants. J Glaucoma, 2011, 20: 7-9.
  • 7Scuderi GL, Cascone NC, Regine F, et al. Validity and limits of the rebound tonometer (ICare (R)): clinical study. Eur J Ophthal- mol, 2011, 21: 251-257.
  • 8Dekking HM, Coster HD. Dynamic tonometry. Ophthalmologica, 1967, 154: 59-74.
  • 9Kontiola A. A new electromechanical method for measuring in- traocular pressure. Doc Ophthalmol, 1996, 93: 265-276.
  • 10Kontiola AI, Goldblum D, Mittag T, et al. The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res, 2001, 73: 781-785.

二级参考文献10

共引文献9

同被引文献11

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部