Lenstar与A超测量前房深度及晶状体厚度的比较 被引量:5

Comparision of Lenstar and A-scan ultrasound biometry in the measurement of anterior chamber depth and lens thickness
下载PDF
导出
摘要 目的比较Lenstar LS900光学生物测量仪与A超测量(ACD)、晶状体厚度(LT)的差异,以评价Len-star测量参数的准确性,为其临床应用提供依据。方法对31例(35只眼)白内障患者分别应用Lenstar与A超测量ACD、LT,测量数据之间的差异采用配对t检验,一致性采用Bland-Altamn统计分析。结果 Lenstar测量ACD为(3.047±0.384)mm,A超为(2.853±0.397)mm,差异有统计学意义(t=2.242,P<0.05),一致性较差。测量LTLenstar为(4.546±0.370)mm,A超为(4.631±0.411)mm,差异无统计学意义(t=0.912,P﹥0.05),一致性较差。结论 Lenstar与A超的测量结果存在一定差异,在临床中应注意仪器之间的差异。 Objective Compare the differences between Lenstar and A-scan ultrasound biometry for measuring the anterior chamber depth and lens thickness in order to evaluate the accuracy of those two instruments. Methods Totally 35 eyes of 31 cataract patients were measured with both Lenstar and A-scan ultrasound biometry separately. Date were analyzed using a paired samples t test and Blant-Altman plots. Results The measurements of anterior chamber depth with Lenstar was ( 3. 047 ± 0. 384) mm. A-scan ultrasound biometry was ( 2. 853 ± 0. 397 ) ram, the difference was statistically significant ( t = 2.242, P 〈 0.05 ). The measurements of lens thickness with Lenstar was ( 4. 546 ± 0. 370 ) nun, A-scan ultra- sound biometry was ( 4.631 ± 0.411 ) mm, the difference was not statistically significant ( t = 0. 912, P 〉 0. 05 ). The Bland-Ahanm showed that the two decices had nan-comparable results for anterior chamher depth and lens thickness. Condusion Attention should be paid to the differences between Lenstar and A-scan ultrasound biometry in clinical practice. The two device may not be directly substituted for each other.
出处 《临床眼科杂志》 2012年第5期418-420,共3页 Journal of Clinical Ophthalmology
关键词 LENSTAR A超 前房深度 晶状体厚度 Lenstar A-scan ultrasound biometry Anterior chamber depth Lens thickness
  • 相关文献

参考文献12

  • 1Bland JM,Altman IX;. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet, 1986,327 : 307-310.
  • 2黄锦海,陈世豪,温岱宗,王勤美.Biograph/Lenstar与IOLMaster测量眼轴、角膜曲率及前房深度的比较[J].中华眼视光学与视觉科学杂志,2011,13(2):126-130. 被引量:22
  • 3Shammas HI. Intraocular IOL power calculations, mustmted ed Thorofare, NJ: SLACK Incorporated,2004:19.
  • 4Salouti R, Nowroozzadeh MH, Zamani M, et al. Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraoctdar lens power calculation. Optometry, 2011, 82 : 140-147.
  • 5Vetrugno M, Cardascia N, Cardia L. Anterior chamber depth measmed by two methods in myopic and hyperopic phakie IOL implant. Br J Ophthalmol, 2000,84:1113-1136.
  • 6O'Donnell C, Hartwig A, Radhakrishnan H. Correlations between refractive error and biometrie pammetes in human eyes using the LeuStar 900. Contact Lens Anterior Eye,2011,34:26V31.
  • 7Shammas HJ. Intmocular IOL power calculations, Illustrated ed Thorofare, NJ: SLACK Incorpomted ,2004 :17.
  • 8Olsen T. Caleulation of intraocular lens power: a review. Aeta Ophthalmol Stand, 2007,85:472-485.
  • 9Bunkhurst PJ, Wolfsohn JS, Shah S, et al. A new optical low co- herence refleetometry device for ocular biometiy in cataract pa- tients. Br J Ophthalmol, 2009,93:949-953.
  • 10Lehman BM, Bemtsen DA, Bailey MD, et al. Validation of opti- cal coherence tomography - based crystalline lens thickness meas- urements in children. Optom Vis Sci ,2009,86 : 181-187.

共引文献21

同被引文献64

  • 1徐海铭,施明光,张艳玲,马宝峰,郑穗联.OrbscanⅡ系统在测量晶状体厚度、眼轴及前房深度中的应用[J].临床眼科杂志,2004,12(4):297-299. 被引量:3
  • 2葛坚,刘炳乾,高前应.角膜厚度与青光眼危险性的关系[J].眼科,2005,14(3):145-146. 被引量:29
  • 3刘国军,沈宝莲,王海存,康维强,刘泳,刘志宏,仇宜解.硅油填充眼眼轴的超声测量[J].中华超声影像学杂志,2005,14(7):559-560. 被引量:9
  • 4阎启昌,邸悦,董雅娟,张劲松.眼前节图像分析系统与A超测量前房深度及晶状体厚度的比较[J].中国实用眼科杂志,2006,24(6):602-604. 被引量:7
  • 5吴敏,叶剑,孙强,翟军印,曹书,喻箭.前节OCT与A超测量前房深度的比较[J].国际眼科杂志,2007,7(5):1341-1342. 被引量:8
  • 6O,Donnell C,Hartwig A, Radhakrishnan H. Correlations between refractive error and biometric parameters in human eyes using the Lenstar 900[J]. British Contact Lens Association,2010,34(1): 26-31.
  • 7Salouti R,Nowroozzadeh MH,Zamani M,et al. Comparison of the ultrasonographic method with partial coherence interferometry methods for intraocular lens power calculation[J]. American Op- tometric Association, 2011,82(3):140-147.
  • 8Rohrer K,Frueh BE,Wahi R. Comparison and evaluation of ocular biometry using a new noncontact opitcal low coherence refleetometer[J]. Ophthalmology, 2009,116(11):2087-2092.
  • 9Bunkhurst PJ,Wolfsohn JS,Shah S.A new optical low coherence reflectometry device for ocular biometer in cataract patients[J].Br J Ophthalmol, 2009,93(7):949-953.
  • 10Bjelos Roncevic M,Busic M,Cima I,et al. Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer[J]. Graefes Arch Clin Exp Ophthalmol,2011,249(1):83-87.

引证文献5

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部