摘要
目的评估可折叠房角支撑型人工晶状体(IOL)植入术矫正高度近视的安全性、有效性、可预测性及稳定性。方法收集2010年4月至2011年7月在我院接受可折叠房角支撑型IOL(CACHET房角支撑型IOL,美国ALCON公司)植入术的高度近视患者共15人(29只眼),术后随访1年。术前平均球镜度(-12.08±2.44)D(范围:-8.75~-16.50D)。术后1个月、3个月、6个月、1年检查裸眼视力(UCVA)、最佳矫正视力(BSCVA)、角膜内皮密度、IOL前房稳定性及手术并发症,评估手术的安全性、有效性、可预测性及稳定性。对角膜内皮细胞密度采用“检验进行统计学分析。结果术后1年,BSCVA≥0.6者为100%,BSCVA≥1.0者89.7%(26/29);BSCVA较术前下降者为0眼;较术前无变化的占62.1%(18/29);较术前提高1行的占31.0%(9/29);较术前提高2行的占3.4%(1/29);较术前提高大于2行的占3.4%(1/29);手术安全指数为1.13(1.08/0.96)。术后1年,100%术眼UCVA/〉0.5,UCVA≥1.0者58.6%(17/29),UCVA/〉0.8者86.2%(25/29),UCVA≥0.6者93.1%(27/29);手术有效指数为1.03(0.99/0.96)。术后1年,平均球镜度为(-0.23±0.25)D;术后实际屈光度与术前目标屈光度差异±0.50D以内者占100%;平均角膜内皮细胞密度变化为-0.23%±4.80%。术后并发症主要为术后早期眼压升高,占17.2%,仅见于术后当日。眩光及光晕主要出现于术后1个月内,分别占6.9%(2/29)、13.8%(4/29),不影响夜间驾车。结论经过1年的临床观察,显示可折叠房角支撑型人工晶状体植入术矫正高度的安全性、有效性、可预测性强,IOL前房稳定性良好,手术并发症少。由于术后随访时间有限,更长期的临床随访观察很有必要。
Objective To assess the safety, efficacy, predictability and stability of the phakic foldable angle-supported intraocular lens (IOL) (CACHET IOL, Aleon Inc. , USA. ) for correction of high- to-extremely high myopia in adults. Methods In this prospective study, 29 eyes of 15 patients were implanted with CACHET phakic angle-supported IOL and followed for 1 year postoperatively. Preoperative manifest refractive sphere was - 12. 08 ± 2. 44 diopters ( D). Visual acuity ( uncorrected visual acuity, UCVA ; best spectacle-corrected visual acuity, BSCVA ) , predictability and stability of manifest refraction spherical equivalent (MRSE) , adverse events, and endothelial cell density were analyzed primarily during 1 -year of follow-up. Results After 1 year of follow-up, a BSCVA of 0. 6 or better was achieved by 100% of eyes; 9.7% (26/29) had a BSCVA of 1.0 or better; no eyes lost≥1 line BSCVA, 62. 1% (18/29) had no change in lines of BSCVA, 31.0% (9/29) gained 1 line of BSCVA, 3.4% (1/29) gained 2 line of BSCVA, and 3.4% (1/29) gained more than 2 line of BSCVA; the safety index was 1.13 ( 1.08/0. 96) ; UCVA of 1.0 or better was achieved by 58.6% ( 17/29), 100% had an UCVA of 0. 5 or better, 86. 2% (25/29) of eyes achieved a UCVA of 0. 8 or better, and 93. 1% (27/29) of eyes achieved a UCVA of O. 6 or better; the efficacy index was 1.03 (0. 99/0. 96). The overall mean percentage change in endothelial cell density 1 year after surgery was -0. 23% ±4. 80%. Five eyes ( 17.2% ) had increased intraoeular pressure (IOP) on the day of surgery. No pupil ovalization, pupillary block, or retinal detachment events were observed. Glare(6. 9%, 2/29) and halo(13.8% ,4/29) mainly appeared in the first month after surgery, and had no influence on driving in night. Conclusions After 1 year of follow-up, the implantation of phakic foldable angle-supported IOL was proved to be safe, effective and predictable with minimal complications in patients with high myopia. Due to the limitation of visiting time, long-term of clinical investigation is necessary to verify the safety and efficacy of this IOL.
出处
《中华眼科杂志》
CAS
CSCD
北大核心
2012年第9期804-810,共7页
Chinese Journal of Ophthalmology