期刊文献+

床边检测血糖仪与生化仪检测血糖相关性分析 被引量:4

Correlation analysis of the POCT blood glucose meter and biochemical analyzer
原文传递
导出
摘要 目的:对床边检测血糖仪测定结果与全自动生化分析仪血糖测定结果进行对比,评价其相关性。方法:采用40份不同浓度血糖的EDTA-K2抗凝全血,在POCT血糖仪上测定三次,取均值,同时离心取血浆在全自动生化分析仪上测定三次,取均值进行比较分析。结果:POCT血糖仪检测结果比全自动生化分析仪检测血糖结果稍低,但在医学决定水平浓度的系统误差<10%,能够被临床所接受。结论:POCT血糖仪与全自动生化仪测定血糖的相关性尚好,可以用于糖尿病患者血糖监测或过筛,但是不能作为糖尿病的确诊试验。 Objective:To compare the blood glucose results tested by POCT glucose meter and biochemical analyzer,and to evaluate their interrelation.Methods:The venous blood specimens from 40 patients were detected with POCT glucose meter,while plasma glucose was detected with biochemical analyzer for three times respectively.The venous blood was anti-coagulated with EDTA-K2,and centrifuged before testing by biochemical analyzer.The results of three tests were averaged and interrelation analysis was performed between two methods.Results:The glucose results of POCT were slightly lower than that derived from biochemical analyzer,but system error 10%,which can be accepted by clinical work.Conclusion:The results of blood glucose acquired from POCT glucose testing method were were close to those from biochemical analyzer,showing that POCT glucose meter can be used as the diabetic screening and blood glucose monitoring for diabetic.But it can not be used in confirmation test for diabetic.
出处 《中国卫生检验杂志》 北大核心 2012年第9期2130-2131,2133,共3页 Chinese Journal of Health Laboratory Technology
关键词 POCT血糖仪 生化分析仪 POCT blood glucose meters Biochemical analyzer
  • 相关文献

参考文献6

  • 1美国床边检验(POCT)介绍[J].上海医学检验杂志,1999,14(3):132-133. 被引量:22
  • 2National Committee for Clinical Laboratory Standards. C50 - A2 Paint-of- care blood glucose testing in acute and chronic care facilities; approved guideline second edition[ S]. Wayne, PA : NCCLS, 2002.
  • 3In vitro diagnostic test systems. Requirements for blood - glucose mon- intoring systems for self - testing in managing diabetes mellitus. Inter- national standard[ Sq, ISO 15197 : 2003 (E).
  • 4Method Comparison and Bias Estimation Using Patient Samples, ap- proved guideline - second edition[ Sq. Clinical And Laboratory Stand- ards Institute, 22 (19) : EP9 -A2.
  • 5唐立萍,居漪,欧元祝,王美娟,张瑞镐.POCT血糖仪的性能分析[J].检验医学,2010,25(1):13-16. 被引量:32
  • 6冯仁丰.美国医院内非检验科进行葡萄糖POCT的管理要求介绍[J].上海医学检验杂志,1999,14(3):139-140. 被引量:52

二级参考文献7

  • 1American Diabetes Association. Consensus statement on self-monitoring of blood glucose [ J ]. Diabetes Care, 1987,10( 1 ) :95-99.
  • 2American Diabetes Association. Bedside blood glucose monitoring in hospitals [ J ]. Diabetes Care, 2004, 27 (Suppl 1 ) :S104.
  • 3G13/T19634-2005.体外诊断检验系统自测用血糖监测系统通用技术条件[S].
  • 4Bastanhagh MH, Shirvan AR, Heshmat R, et al. Evaluation of the efficacy of blood glucose home monitoring devices [ J]. Med Sci Monit, 2007, 13 (3) :PI1-PI6.
  • 5Skeie S, Thue G, Nerhus K, et al. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician [J]. Clin Chem, 2002, 48(7) : 994-1003.
  • 6Solnica B, Naskalski JW, Sieradzki J. Analytical performance of glucometers used for routine glucose self- monitoring of diabetic patients [ J ]. Clin Chim Acta, 2003,331 (1-2) :29-35.
  • 7Stahl M, Brandslund I. Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus[J]. Scand J Clin Lab Invest ,2003, 63(6) : 431-440.