摘要 在X射线荧光分析中,使用滤光片可有效降低原级光谱本底对待测元素的干扰,提高分析灵敏度。本文采用MCNP5方法仿真X光管滤光片对原级谱线的影响,比较了滤光片对不同能量电子束产生的X射线谱的差异,分析了不同滤光片材料及厚度与X射线原级谱衰减的关系,得到不同滤光片对X射线谱的衰减规律。 In X-ray fluorescence (XRF) analysis, the use of filter reduces the background interference effectively, and improves the analysis sensitivity. In this paper, the MCNP5 code is used to simulate the filter effects in X-ray tube XRF. The XRF spectra by 140 keV electron beams, with filters of different materials and thicknesses, are compared. Their attenuation rates, i.e. background reduction from the original XRF spectra, are analyzed.
作者 卢艳 黄宁
出处 《核技术》 CAS CSCD 北大核心 2012年第10期751-754,共4页 Nuclear Techniques
关键词 X光管 X射线谱 滤光片效应 MCNP5仿真 X-ray tube, X-ray spectra, Filter effect, MCNP5 simulation
  • 相关文献

参考文献7

  • 1Taleei R, Shahriari M. Monte Carlo simulation of X-rayspectra and evaluation of filter effect using MCNP4C andFLUKA code[J]. Applied Radiation and Isotopes, 2009,2(67): 266-271.
  • 2Ay M R,Shahriari M. Monte Carlo simulation of X-rayspectra in diagnostic radiology and mammography usingMCNP4C[J]. Physics in Medicine and Biology, 2004,21(49): 4897-4917.
  • 3Querol A, Gallardo S. Uncertainty analysis in X-rayspectra simulation:effect of geometric tube features[C].International Conference on Biomedical Engineering andInformatics, 2011.
  • 4张静,谢亚宁,侯凯,胡天斗,刘涛.X射线吸收光谱技术中的滤光片研制[J].核技术,2004,27(7):497-500. 被引量:2
  • 5贾鹏翔,包尚联.利用Monte Carlo方法研究X射线球管阳极靶角度对输出光谱的影响[A].中国仪器仪表学会医疗仪器分会第四次全国会员代表大会暨2009年学术年会[C]. 2009.
  • 6谈春明.X射线荧光分析原级能谱分布的MCNP模拟[J].原子能科学技术,2010,44(B09):496-499. 被引量:9
  • 7Hubbell J H, Seltzer S M. Tables of X-ray mass attenua-tion coefficients and mass energy-absorption coefficientsfrom 1 keV to 20 MeV for elements Z=1 to 92 and 48additional substances of dosimetric interest[R]. NISTIR5632, National Institute of Standards and Technology,Gaithersburg, MD, USA.

二级参考文献7

  • 1纪新明,王建业,贾文红,朱节清,黄宜平.X射线荧光分析中X射线管原级能谱分布的测定[J].光学学报,2006,26(4):634-638. 被引量:10
  • 2杨明太.X射线荧光光谱仪的现状[J].核电子学与探测技术,2006,26(6):1025-1029. 被引量:37
  • 3JUDITH F B.MCNP-A general Monte Carlo N-particle transport code,version 4C,LA-13709-M[R].US:Los Alamos National Laboratory,2000.
  • 4PELLA P A,FENG L Y,SMALL J A.An analytical algorithm for calculation of spectral distributions of X-ray tubes for quantitative X-ray fluorescence analysis[J].X-ray Spectrometry,1985,14(3):125-135.
  • 5Stern E A, Heald S M. Rev Sci Instrum, 1979, 50(12):1579-1582
  • 6Gullikson E M. Mass Absorption Coefficient, X-ray Data Booklet, Lawrence Berkeley National Laboratory, Univercsity of California Berkeley, CA 94720, January 2001
  • 7Bunker G., Optimizing X-ray Filters, Biophysics Collaborative Access Team Basic Techniques For EXAFS,Revision date 6/25/94