期刊文献+

声辐射力脉冲成像技术对正常肾脏的定量研究 被引量:28

A quantitative study of acoustic radiation force impulse imaging in normal kidney
原文传递
导出
摘要 目的确定正常肾脏各部位弹性系数参考值范围并探讨其影响因素。方法选取380名健康体检者共760个正常肾脏,应用声辐射力脉冲成像技术(ARFI)于不同取样深度下测量正常肾脏皮质、髓质、肾窦剪切波传播速度(Vs)并进行比较,分析受试者性别、年龄、身高、体重、取样深度及取样角度对Vs值的影响。结果垂直取样时正常肾脏皮质、髓质、肾窦Vs值参考值范围分别为(3.33±0.56)、(2.29±0.20)、(1.06±0.19)m/s,且差异有统计学意义(F=29451.58,P<0.05);不同侧别、性别、年龄、体重指数(BMI)、健康体检者正常肾脏皮质、髓质、肾窦同一部位Vs值差异均无统计学意义;不同取样深度测量正常肾脏肾皮质Vs值差异也均无统计学意义;取样线与垂直方向呈30°角时测得正常肾脏皮质、髓质、肾窦Vs值分别为(3.01±0.51)、(2.03±0.21)、(0.96±0.15)m/s,各部位比较差异有统计学意义(F=16033.17,P<0.05);与垂直取样时比较差异也均有统计学意义(t值分别为7.588、2.821、3.946,P均<0.05)。结论 ARFI技术可较客观地测量正常肾脏不同部位Vs值,且影响因素较少,可定量评价正常肾脏不同部位组织弹性顺应性。 Objective To define the reference range of elastic coefficient of different parts of normal kidneys and to investigate the influencing factors.Methods Totally 760 kidneys from 380 healthy people were selected in this study.The shear wave velocities(Vs)of normal renal cortex,renal medulla and renal sinus were measured using acoustic radiation force impulse imaging(ARFI)at different sampling depth.The recorded Vs were compared.The influence of gender,age,height,weight,sampling depth and sampling angle on Vs was analyzed.Results The Vs reference range of normal renal cortex,medulla and sinus at vertical sampling angle was(3.33±0.56),(2.29±0.20),and(1.06±0.19)m/s,respectively.The difference was statistically significant(F=29 451.58,P0.05).When the effects of lateral position,gender,age and body mass index(BMI)on Vs were compared,there was no statistical significance for Vs obtained from the same parts of renal cortex,medulla and sinus of each individual.The Vs of renal cortex was no statistically significant at different sampling depth.When the sampling line was at 30°angle to the vertical direction,the Vs of normal renal cortex,medulla and sinus was(3.01±0.51),(2.03±0.21),and(0.96±0.15)m/s,respectively.The difference was statistically significant(F=16 033.17,P0.05).The Vs obtained from difference parts of kidney at vertical sampling angle were significantly different to the corresponding results obtained when the sampling line was at 30°angle(renal cortex:t=7.588,P0.05,renal medulla:t=2.821,P0.05,renal sinus:t=3.946,P0.05).Conclusions The ARFI technology can objectively measure the Vs reference range of the different parts of normal kidneys,and meanwhile the Vs value was least influenced by other factors.So it can quantitatively evaluate tissue elasticity of different parts of normal kidneys.
出处 《中华医学超声杂志(电子版)》 2012年第10期53-56,共4页 Chinese Journal of Medical Ultrasound(Electronic Edition)
关键词 正常肾脏 成像技术 辐射力 脉冲 参考值范围 健康体检者 深度测量 影响因素 Elasticity imaging technigues Kidney cortex Kidney medulla
  • 相关文献

参考文献15

  • 1王正滨,孟冬梅.声脉冲辐射力成像技术在肾脏疾病诊断中的应用与研究进展[J].中华医学超声杂志(电子版),2012,9(5):9-11. 被引量:8
  • 2Thomas A,Fischer T,Frey H. Real-time elastography an advancedmethod of ultrasound:first results in 108 patients with breast lesions[J].Ultrasound in Obstetrics and Gynecology,2006,(03):335-340.
  • 3Miyanaya N,Akaza H,Yamakawa M. Tissue elasticity imagingfor diagnosis of prostate cncer:a preliminary report[J].International Journal of Urology,2006,(12):1514-1518.
  • 4Cochlin DL,Ganatra RH,Grifiths DF. Elastogrphy in the detection ofprostatic cancer[J].Clinical Radiology,2002,(22):1014-1020.
  • 5相蕾,黄慧,吕泽平,万钢,胡才友,刘红,邓宏明,李必迅,宋玉国,齐科研,唐雷,孙亮,史晓红,杨泽.高血压、肥胖、血脂紊乱和糖耐量异常在不同年龄段人群中分布的基线调查 社区人群心脑血管事件前瞻性研究系列之五[J].中国老年保健医学,2009,7(1):22-26. 被引量:16
  • 6王正滨;唐杰;杨斌.泌尿生殖系统疾病超声诊断与鉴别诊断学[M]北京:人民卫生出版社,201076-82.
  • 7Ophir J,Cespedes I,Ponnekanti H. Elastography:a quantitativemethod for imaging the elasticity of biological tissues[J].Ultrasonic Imaging,1991,(02):111-134.
  • 8Frey H. Real-time elastography A new ultrasound procedure foe thereconstruction of tissue elasticity[J].Radiology,2003,(10):850-855.
  • 9Fahey BJ,Palmeri ML,Trahey GE. Frame rate consideration forreal-time abdominal acoustic radiation force impulse imaging[J].Ultrasonic Imaging,2006,(04):193-210.
  • 10Gallotti A,D′Onofrio D,Pozzi Mucelli R. Acoustic radiation forceimpulse( ARFI )technique in ultrasound with virtual touch tissuequantification of upper abdomen[J].La Radiologia Medica,2009,(06):889-897.

二级参考文献49

共引文献75

同被引文献301

引证文献28

二级引证文献89

相关作者

内容加载中请稍等...