摘要 A simple and efficient method using enzyme immobilized microfluidic channel as open tubular microreactor was designed for amperometric detection of glucose. The microreactor was composed of a polydimethylsilicone/ glass hybrid device with three reservoirs, a cooling cave and a 6 cm capillary with a sampling fracture as micro-channel. The microchannel was further modified by thermal polymerization, followed by covalently attaching with glucose oxidase. Through fracture sampling and electrochromatography separation, the production via enzymatic reaction was determinated by Pt electrode at the end of capillary. The linear range for the detection of glucose was 0.05--7.5 mmol·L-1 with detection limit of 23μmol.L-1 The inter-and intra-chip reproducibilities for determination of 2.5 mmol-L-1 glucose were 98.5% (n=5) and 96.0% (n=5), respectively. With the advantage of flexible assembly, rapid efficiency, good stability and low-cost, this microreactor provided a potential platform for estab- lishing a portable enzyme-based chemical detection system in practical application. A simple and efficient method using enzyme immobilized microfluidic channel as open tubular microreactor was designed for amperometric detection of glucose. The microreactor was composed of a polydimethylsilicone/ glass hybrid device with three reservoirs, a cooling cave and a 6 cm capillary with a sampling fracture as micro-channel. The microchannel was further modified by thermal polymerization, followed by covalently attaching with glucose oxidase. Through fracture sampling and electrochromatography separation, the production via enzymatic reaction was determinated by Pt electrode at the end of capillary. The linear range for the detection of glucose was 0.05--7.5 mmol·L-1 with detection limit of 23μmol.L-1 The inter-and intra-chip reproducibilities for determination of 2.5 mmol-L-1 glucose were 98.5% (n=5) and 96.0% (n=5), respectively. With the advantage of flexible assembly, rapid efficiency, good stability and low-cost, this microreactor provided a potential platform for estab- lishing a portable enzyme-based chemical detection system in practical application.
出处 《Chinese Journal of Chemistry》 SCIE CAS CSCD 2012年第9期2145-2150,共6页 中国化学(英文版)
关键词 MICROREACTOR amperometric detection GLUCOSE enzymatic catalysis microfluidic channel microreactor, amperometric detection, glucose, enzymatic catalysis, microfluidic channel
  • 相关文献

参考文献28

  • 1Kovarik, M. L.; Gach, P. C.; Ornoff, D. M.; Wang, Y. L.; Balowski, J.; Farrag, L.; Allbritton, N. L. Anal. Chem. 2012, 84, 516.
  • 2Marcus, J. S.; Anderson, W. F.; Quake, S. R. Anal. Chem. 2006, 78, 956.
  • 3Herr, A. E.; Hatch, A. V.; Throckmorton, D. J.; Tran, H. M.; Bren- nan, J. S.; Giannobile, W. V.; Singh, A. K, PNAS 2007, 104, 5268.
  • 4Chen, J.; Li, J.; Sun, Y. Lab. Chip 2012, 12, 1753.
  • 5Wu, H. L.; Yang, P. Y.; Fan, G. R.; Tian, Y. P.; Lu, H. J.; Jm, H. Chin. J. Chem. 2006, 24, 903.
  • 6Chen, Y. J.; Roller, E. E.; Huang, X. H. Lab. Chip 2010, 10, 1153.
  • 7Aravamudhan, S.; Kumar, A.; Mohapatra, S.; Bhansali, S. Biosens. Bioelectron. 2007, 22, 2289.
  • 8Tennico, Y. H.; Hutanu, D.; Koesdjojo, M. T. Bartel, C. M.; Remcho, V. T. Anal. Chem. 2010, 82, 5591.
  • 9Mersal, G. A. M.; Bilitewski, U. Electrophoresis 2005, 26, 2303.
  • 10He, P.; Greenway, G.; Haswell, S. J. Microfluid Nanofluid 2010, 8, 565.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部