眼眶部螺旋CT检查螺距选择与辐射剂量的相关性探讨

The Research of the Relativity Between the Pitch and Radiation Dose at Orbital Helical CT
原文传递
导出
摘要 目的探讨眼眶部螺旋CT检查螺距变化与图像质量、辐射剂量的相关性。方法分析80例眼眶部螺旋CT扫描的断面图像和重组图像,螺距分别为0.75、1.5各40例。同时记录容积CT剂量指数(CTDIvol)和剂量长度乘积(DLP)。图像质量从影像层次、背景噪声、解剖结构及能否满足诊断要求等方面进行综合评价;图像资料采用χ2检验。结果螺距0.75、1.5的CTDIvol和DLP分别为23.01 mGy、131.18 mGy.cm和11.51 mGy、69.04mGy.cm,呈负相关;两组断面图像和重组图像的质量间差异无统计学意义(P>0.05)。结论眼眶部螺旋CT检查螺距设置1.5时,既能获得良好的图像质量,又能有效地降低辐射剂量。 Objective To investingate the relativity of orbital helical CT pitch changes,image quality and radiation dose.Methods 80 cases of orbital spiral CT scan of the sectional images and reconstructive images were analyzed.The pitch were respectively 0.75,1.5 with 40 cases in each.At the same time CT dose index(CTDIvol) and the dose length product(DLP) were recorded.Image quality was evaluated according to its low-density resolution,background noise,anatomic structure and whether it could meet the diagnosis requirements.The image data were analyzed by χ2 test.Results The CTDIovl and DLP of Pitch 0.75,1.5 CTDIovl and DLP were 23.01 mGy、131.18 mGy·cm and 11.51 mGy、69.04 mGy·cm respectively,in a negative relationship;two sets of image quality difference had no statistical significance(P〉 0.05).Conclusion Orbital helical CT with 1.5 pitch,not only obtain good image quality,but also can effectively reduce radiation dose.
出处 《临床放射学杂志》 CSCD 北大核心 2012年第10期1473-1475,共3页 Journal of Clinical Radiology
关键词 眼眶 螺距 体层摄影术 X线计算机 辐射剂量 Orbital Pitch Tomography X-ray computed Radiation dose
  • 相关文献

参考文献7

二级参考文献25

  • 1郑钧正,岳保荣,李述唐,王琪,张建峰.我国“九五”期间X射线诊断的医疗照射频率水平[J].中华放射医学与防护杂志,2001(S1):20-24. 被引量:69
  • 2华建明,陈根松,丁信法,向可伟,章伟敏.颞骨低剂量CT扫描的应用研究[J].实用放射学杂志,2006,22(2):170-172. 被引量:14
  • 3王卉,郑晓华,金朝林.眼眶低剂量CT扫描的应用价值[J].中华放射医学与防护杂志,2007,27(2):198-199. 被引量:13
  • 4赵勇,李品江.CT噪声和伪影对成像质量的影响及控制[J].中华放射学杂志,1997,31(1):56-58. 被引量:11
  • 5Haaga JR.Radiation dose management:weighing risk versus benefit AJR,2001,177:289-291.
  • 6Slovis TL.CT and computed radiography:the pictures are great,but is the radiation dose greater than required? AJR,2002,179:39-41.
  • 7Wilting J,Zwartkruis A,van Leeuwen MS.A rational approach to dose reduction in CT; individualized scan protocols.Eur Radiol,2001,11:2627-2632.
  • 8Paulson EK,Weaver C,Ho LM,et al Conventional and reduced radiation dose of 16-DMCT for detection of nephrolithiasis and ureterolithiasis.AJR,2008,190:151 -157.
  • 9Hanai K,Horiuchi T,Sekiguchi J,et al.Computer-simulation technique for low dose computed tomographic screening.J Comput Assist Tomogr,2006,30:955-961.
  • 10Bax JJ,Schuijf JD.Which patients should be referred for noninvasive angiography with multi-slice CT? Int J Cardiol,2007,114-113.

共引文献62

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部