期刊文献+

基于量子级联激光器光腔衰荡光谱技术的痕量氨气检测 被引量:7

Cavity ring-down spectroscopy for trace ammonia detection
下载PDF
导出
摘要 采用基于红外脉冲量子级联激光器的高灵敏光腔衰荡光谱(Cavity ring-down spectroscopy)技术建立了氨气在1027 cm-1(9.7μm)附近的吸收光谱痕量检测实验装置.使用该装置对人体口腔呼出气体以及超净实验室环境中氨气成份进行了测量和分析,氨气的检测灵敏度达到10 ppb.模拟分析了激光器光谱线宽对测量结果和系统检测灵敏度的影响,随着激光器光谱线宽的增加,测量结果偏低,系统检测灵敏度下降.为了减小光源线宽对测量结果的影响,通过修正曲线对测量数据进行修正,得到了较好的结果. A cavity ring-down spectroscopy apparatus employing a pulsed quantum cascade laser operating near 1027 cm^-1 (9.7 μm) is investigated for trace ammonia detection. The trace ammonia in human breath and clean room atmosphere are measured and analyzed. A sensitivity of 10 ppb has been demonstrated for ammonia detection. The effects of the spectral line-width on the measurement accuracy and sensitivity of the system are analyzed. The measurement accuracy and sensitivity of the apparatus decrease with the increasing laser spectral line-width. The measurement results can be corrected by taking into account of the line-width effect in data analysis.
出处 《红外与毫米波学报》 SCIE EI CAS CSCD 北大核心 2012年第5期431-436,共6页 Journal of Infrared and Millimeter Waves
基金 国家自然科学基金项目(60878038)~~
关键词 光腔衰荡光谱 量子级联激光器 氨气 线宽 cavity ring-down spectroscopy quantum cascade laser ammonia line-width
  • 相关文献

参考文献11

  • 1O'KEEFE A, DEACON D A G. Cavity ring-down optical spectrometer for absorption measurement using pulsed laser sources[J]. Rev. Sci. Instrum. , 1988,59 ( 12).:2544 -2561.
  • 2ROMANNI D, KACHANOV A A, SADEGHI N, et al. CW cavity ring down spectroscopy [ J ]. Chem. Phys. Lett., 1997,264:316 - 322.
  • 3DEBECKER I, MOHAMED A K, ROMANIN1 D. High- speed cavity ringdown spectroscopy with increased spectral resolution by simultaneous laser and cavity tuning[ J]. Opt. Express, 2005,13 ( 8 ) : 2906 - 2915.
  • 4SCHOCKER A, HOINGHAUS K K, BROCKHINKE A. Quantitative determination of combustion intermediates with cavity ring-down spectroscopy: systematic study in propene flames near the soot-formation limit[ J]. Appl. Opt. ,2005, 44 ( 31 ) : 6660 - 6672.
  • 5RICHMAN B A, KACHANOV A A, PALDUS B A. Novel detection of aerosols: combined cavity ring-down and fluo- rescence spectroscopy [ J ]. Opt. Express, 2005,13 ( 9 ) : 3376 - 3387.
  • 6WANG Chu-Ji, SAHAY P. Breath Analysis Using Laser Spectroscopic Technique: Breath Biomarkers, Spectral Fin- gerints, and Detection Limits [ J ]. Sensors,2009,9 : 8230 - 8262.
  • 7NARASIMHAN L R, GOODMAN W, PATEL C K N. Cor- relation of breath ammonia with blood urea nitrogen and cre- atinine during hemodialysis [ J]. PANS. 2001,98 ( 8 ) :4617 - 4621.
  • 8ROTHMAN S, JACQUEMART D, BARBE A, et al. The HITRAN 2004 molecular spectroscopic database [ J ]. J. Quantitative Spectroscopy& Radiative Transfer. 2005,96 : 139 - 204.
  • 9朱方强,姜开利,何秋荣,董国轩,马辉,陈瓞延.腔衰荡吸收测量中光源线宽影响研究[J].Chinese Journal of Chemical Physics,1998,11(1):26-30. 被引量:5
  • 10ZALICKI P, ZARE R N. Cavity ring-down spectroscopy for quantitative absorption measurements [ J ]. J. Chem. Phys. , 1995,102 (7) :2708 - 2717.