期刊文献+

智能型眼压仪的研制及临床应用 被引量:1

The manufacture of intelligent tonometer and its application
下载PDF
导出
摘要 目的 研制一种自动化程度高、成本低、易普及的智能型眼压仪。方法 采用压陷式测量原理,利用传感技术、计算机技术、电子技术研制成眼压仪。并对其准确性进行临床观察,与Schiotz眼压计、CT-60型非接触眼压仪进行比较。结果智能型眼压仪与Schiotz眼压计、CT-60型非接触眼压仪的准确性天显著性差异,相关性好。智能型眼压便灵敏性、精确度优于Schiotz眼压计。成本远比CT-60型非接触眼压仅低。结论 智能型眼压仪经临床验证各项性能指标均符合设计要求,是一种实用的眼压测量仪器。 Objective To manufacture a kind of intelligent tonometer that is more automatic, cheaper and easier to be used. Methods We designed a tonometer by the compressed measurement principle, sensor and computer technique,anal- ysed the accurate with a quantity of clinical investigations, and compared it to Schiotz tonometer, and CT-60 non-contact tonometer. Results The accurate of this tonometer was same as that of Schiotz and CT-60 non-contact tonometer, and it has a good relatives with the two kinds above. But the sensitivity and accurate of intelligent tonometer were better than that of Schiotz tonometer; the cost was lower than that of CT-60 non-contact tonometer. Conclusion The intelligent tonometer is a suitable and practical instrument in clinical practice.
机构地区 吉林省医院眼科
出处 《中国实验诊断学》 2000年第4期164-165,共2页 Chinese Journal of Laboratory Diagnosis
基金 吉林省科委资助项目!(963017)
关键词 传感器 眼压检测 智能型眼压仪 Sensor: Computer: Detection of eyeball tens
  • 相关文献

参考文献3

  • 1周毅丁 朱文荣.Perkins压平眼压计与Schiotz压陷眼压计测量正常人及青光眼眼压对比研究[J].实用眼科杂志,1988,6(6):345-345.
  • 2迟卉,王尔光.Schitz压陷眼压计的临床应用估价[J].实用眼科杂志,1990,8(12):732-733. 被引量:3
  • 3施益浩.非接触眼压计正常人眼眼压测定[J].中华眼科杂志,1981,17(14):211-211.

共引文献2

同被引文献20

  • 1孙奎.眼压测量探讨[J].中国眼镜科技杂志,2006(9):121-121. 被引量:1
  • 2方春庭,俞振飞,金成鹏.一种新型凸面眼压计的设计原理及研制[J].临床眼科杂志,2007,15(4):342-344. 被引量:1
  • 3中国人正常眼压的测定[J]山西医学院学报,1961(04).
  • 4Peter W Dekker,Yves C A Robert,Hartmut Kanngiesser,Peter Pirani,Beatrix Entenmann.Principles of contact lens tonometry[J]. International Ophthalmology . 1998 (2)
  • 5Friedenwald JS.Standardization of Tonometers:Decennial Report by the Committee on Standardization of Tonometers. Am Acad Ophthalmol Otolaryngol . 1954
  • 6Perkins,E.S.Hand-held applanation tonometer. British Journal of Ophthalmology . 1965
  • 7Mackay,R. S.,Marg,E.Fast, automatic, electronic tonometers based on an exact theory. Acta ophthal. (Kbh.) . 1959
  • 8Marg,E.,Mackay,R. S.,Oechsli,R.Corneal bending and buckling in tonometry. A.M.A. Arch. Ophthal . 1961
  • 9B.Grolman.A New Tonometer System. American J.Optometry and Archives of AmericanAcademy of Optometry . 1972
  • 10Shields MB.The non-contact tonometer.Its value and limitations. Survey of Ophthalmology . 1980