管电压100kV下双源CT冠状动脉成像的可行性 被引量:12

Feasibility of Dual-source CT Coronary Artery Angiography with 100 kV Tube Voltage
下载PDF
导出
摘要 目的探讨双源CT100kV管电压冠状动脉成像的可行性。资料与方法60例行双源CT冠状动脉造影患者,体重指数<25.0kg/m2、心率<75次/min、窦性心律且规整,随机分成100kV、362mA及120kV、400mA两组,每组30例。比较两组的影像质量、平均CT值、图像噪声、信噪比及辐射剂量。结果双源CT冠状动脉造影显示两组影像质量评分分别为(3.50±0.61)分和(3.30±0.86)分,差异无统计学意义(P>0.05);100kV组平均CT值为(646.50±127.15)Hu,明显高于120kV组的(471.35±88.13)Hu(P<0.05);100kV组图像噪声高于120kV组(P<0.05),但两组信噪比差异无统计学意义(P>0.05);120kV、100kV组有效辐射剂量分别为(10.39±0.36)mSv和(7.29±0.36)mSv,差异有统计学意义(P<0.05)。结论对体重轻、心率不快而律齐者采用管电压100kV行冠状动脉CT血管造影,可在保证图像质量满足诊断的前提下减少辐射剂量。 Purpose To explore the feasibility of dual-source computed tomography coronary angiography(CTCA) at tube voltage of 100 kV.Materials and Methods Dual-source CTCA was performed in 60 consecutive patients with body mass index(BMI) lower than 25.0 kg/m 2 and heart rate less than 75 beats/min.60 cases were randomly divided into two groups.Thirty were examined with a standard protocol(120 kV/400 mA),30 were examined at 100 kV/362 mA.The differences in image quality,mean CT values,image noise,signal-to-noise ratio and radiation dose between two groups were compared.Results The coronary artery image quality scores for group 120 kV and 100 kV were(3.50±0.61) and(3.30±0.86),respectively.There was no statistical difference between the two groups(P 0.05).The mean CT value in 100 kV group [(646.50±127.15) Hu] was statistically higher than 120 kV group [(471.35±88.13)(P0.05).The image noise in 100 kV group was significantly higher than 120 kV group(P0.05).The SNR in 100 kV group was higher than 120 kV group,but there was no statistical difference between the two groups(P0.05).The effective radiation dose was(10.39±0.36) mSv for 120 kV group and(7.29±0.36) mSv for 100 kV group with statistical difference(P0.05).Conclusion In patients with BMI25.0 kg/m 2,resting heart rate 75 beats/min and normal heart rhythm,CTCA at tube voltage of 100 kV can significantly reduce the radiation dose without reducing image quality.
出处 《中国医学影像学杂志》 CSCD 北大核心 2012年第10期734-737,共4页 Chinese Journal of Medical Imaging
基金 云南省教育厅科学研究基金项目(2010Y177) 云南省应用基础研究面上项目(2008c0477m)
关键词 冠状血管造影术 体层摄影术 X线计算机 图像处理 计算机辅助 辐射剂量 管电压 Coronary angiography Tomography X-ray computed Image processing computer-assisted Radiation dosage Tube voltage
  • 相关文献

参考文献3

二级参考文献15

  • 1高建华,孙宪昶,李剑颖,王贵生,李娜,戴汝平.后置滤过器C_2对64层螺旋CT冠状动脉成像质量的影响[J].中国医学影像技术,2007,23(1):73-76. 被引量:31
  • 2Pannu HK, Flohr TG, Corl FM, et al. Current concepts in mutidetector row CT evaluation of the coromary arteries: principles, techniques, and anatomy. Radiographics ,2003,23 : 111-125.
  • 3Tack D, Maertelaer V, C, evenois PA. Dose reduction in muhidetector CT using attenuation-based online tube current modulation. AJR,2003 ,181:331-334.
  • 4Prasad SR, Wittnun C, Shepard JA, et al. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality. AJR,2002,179:461-465.
  • 5Rogers LF. Dose reduction in CT:how low can we go? AJR ,2002, 179:299.
  • 6Haaga JR. Radiation dose management: weighing risk versus benefit AJR,2001,177 :289-291.
  • 7Guidelines for coronary angiography: a report of the American College of Cardiology/American Heart Association Task Force on assessment of diagnostics and therapeutic cardiovascular procedures( subcommittee on coronary angiography). J Am Coll Cardiol, 1987,10 : 935-950.
  • 8WHO. Physical status : the use and interpretation of anthropometry: report of a WHO Expert Committee. WHO technical report series 854. Geneva: World Health Organization, 1995:1-2.
  • 9Slovis TL CT and computed radiography: the pictures are great, but is the radiation dose greater than required? AJR,2002,179: 39-41.
  • 10Mahnken AH, Wildberger JE, Koos R, et al. Multislice spiral computed tomography of the heart: technique, current applications and perspective. Cardiovasc Intervent Radiol,2005 ,28 :388-399.

共引文献151

同被引文献87

  • 1彭晋,张龙江,卢光明.双源双能量CT心肌灌注成像的应用进展[J].医学研究生学报,2011,24(7):775-779. 被引量:6
  • 2王锡明,晁宝婷,武乐斌,张云亭,陈海松,李振家,柳澄.64层螺旋CT心胸联合造影在急性胸痛病因鉴别诊断中的价值[J].中华放射学杂志,2006,40(8):812-814. 被引量:39
  • 3李鹏雨,李坤成,杜祥颖,曹丽珍,刘佳宾,杨延辉,梁志刚,祝小莲,刘建.64层螺旋CT三联检查在急性胸痛诊断中的应用[J].中华放射学杂志,2007,41(10):1032-1035. 被引量:51
  • 4Mayo J, Thakur Y. Pulmonary CT angiography as first- line imaging for PE: image quality and radiation dose considerations. Am J Roentgenol, 2013, 200(3): 522-528.
  • 5European Commission. European guidelines on quality criteria for computed tomography, report EUR-16262. Luxembourg: European Commission, 1999.
  • 6Sommer WH, Schenzle JC, Becker CR, et al. Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique. Invest Radiol, 2010, 45(2): 64-71,.
  • 7Moscariello A, Takx RA, Schoepf U J, et al. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol, 2011, 21(10): 2130-2138.
  • 8Hausleiter J, Bischoff B, Hein F, et al. Feasibility of dual-source cardiac CT angiography with high-pitch scan protocols. J Cardiovasc Comput Tomogr, 2009, 3(4): 236-242.
  • 9Kristiansson M, Holmquist F, Nyman U. Ultralow contrast medium doses at CT to diagnose pulmonary embolism in patients with moderate to severe renal impairment: a feasibility study. Eur Radiol, 2010, 20(6): 1321-1330.
  • 10Nyman U, Bjorkdahl P, Olsson ML, et al. Low-dose radiation with 80-kVp computed tomography to diagnose pulmonary embolism: a feasibility study. Acta Radiol, 2012, 53(9): 1004-1013.

引证文献12

二级引证文献112

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部