摘要
用ICP-OES光谱仪对钽中钙、铝、镁、铌、硅和钨的测定进行了研究和探讨。对样品中元素的分析谱线,仪器分析参数,准确度和精密度进行了研究,确定了最佳实验条件。相对标准偏差不大于6.41%。方法具有良好的精密度和准确度,操作简单、快捷。
Ca, A1, Mg, Nb, Si and W in tantalum samples was determined directly by ICP-OES. The experiment factors including spectra of analyzed element, instrument parameters, accuracy and precision were investigated. The relative standard derivation(RSD)is not more than 6.41%. The method is proved to have good precision and accuracy, and to be simple and rapid.
出处
《湖南有色金属》
CAS
2012年第5期68-69,共2页
Hunan Nonferrous Metals
关键词
钽
ICP-OES
铝
钙
镁
铌
硅
钨
ICP - OE S
Tantalum samples
A1
Ca
M g
Nb
Si
W