RBT与修氏眼压计测量特殊患者眼压的一致性评价

Agreement evaluation of rebound tonometer and Schitz tonometer assessing intraocular pressure values in special cases
下载PDF
导出
摘要 目的探讨回弹式眼压计(rebound tonometer,RBT)测量角膜病变、固视能力差等特殊患者眼压的临床应用价值。方法对同一群特殊患者31例35只眼先后采用RBT和修氏眼压计进行测量,应用配对t检验、相关分析和Bland-Altman分析评价两种测量方法所获结果的一致性。结果 RBT和修氏眼压计测得的眼压分别为(27.4±5.5)mmHg和(29.4±5.8)mmHg(1 mmHg=0.133 kPa),两者呈正相关关系(r=0.952,P=0.000),RBT的测量值略低于修氏眼压计,差异有统计学意义(t=-6.53,P=0.000),一致性欠佳。结论 RBT和修氏眼压计均可应用于其他眼压计测量困难的病例,尽管结果不一定准确,但仍可为临床提供参考。RBT为方便定量测量这些特殊患者的眼压提供了新的途径。 OBJECTIVE To study the clinical value of rebound tonometer(RBT) measuring intraocular pressure(IOP) values of special cases with keratopathy,fixation difficulties and so on.METHODS We used RBT and Schitz tonometer for measuring IOP in a special group of patients(31 cases,35 eyes).The paired t test,correlation analysis and Bland-Altman method were used to evaluate the agreement of two kinds of results.RESULTS The mean IOP of RBT and Schitz tonometer were(27.4±5.5) mmHg and(29.4±5.8) mmHg respectively,and they had a positive correlation(r=0.952,P=0.000).The measured values of RBT were slightly lower than that of Schitz tonometer(t=-6.53,P=0.000),and had poor consistency.CONCLUSIONS RBT and Schitz tonometer could be useful in measuring the IOP in those special cases whose IOP could not be measured by other tonometers.Although the results may not be necessarily accurate,they could also provide references to clinical practice.RBT provides a new way for measuring the IOP of these special patients.
出处 《中国中医眼科杂志》 2012年第5期360-362,共3页 China Journal of Chinese Ophthalmology
关键词 眼压 眼压测量 回弹式眼压计 修氏眼压计 角膜病变 intraocular pressure measurement of IOP rebound tonometer Schitz tonometer keratopathy
  • 相关文献

参考文献9

  • 1陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:216
  • 2甄毅,王涛,王文欣,郑松.国产回弹式眼压计的临床评价[J].眼科新进展,2010,30(6):555-559. 被引量:8
  • 3Kontiola A. A new electromechanical method for measuring intraoc- ular pressure[J].Doc Ophthalmol, 1996-1997,93 (3) : 265-276.
  • 4Kontiola A, Goldblum D, Mittag T, et al. The induction/impact tonometer: a new instrument to measure in traocular pressure in the rat[J]. Exp Eye Res,2001,73 (6) :781-785.
  • 5Flemmons MS, Hsiao YC, Dzau J, et al. Home tonometry for man- agemcnt of pediatric glaucoma[J]. Am J Ophthalmol, 2011,152 (3) : 470-478.
  • 6Flemmons MS, Hsiao YC, Dzau J, et al. Icare rebound tonometry in children with known and suspected glaucoma [J]. J AAPOS, 2011,15 (2) : 153-157.
  • 7Toyhoara K, Inoue K, Wakakura M, et al. Comparison of ICARE rebound tonometer with GAT and non-contact tonometer. Atarashii (Journal of the Eye ), 2007,24 ( 3 ) : 355 -359.
  • 8周和政,江文珊,张绍阳,晏颖,陈云辉,杜伟.ICARE回弹式眼压计与NCT眼压计测量结果的比较[J].国际眼科杂志,2009,9(3):484-485. 被引量:12
  • 9Davies LN, Bartlett H, Mallen EA, et al. Clinical evaluation of re- bound tonometer [J]. Acta Ophthalmol Scand, 2006,84 (2) : 206 - 209.

二级参考文献14

  • 1赵希刚,李坤成,李永忠.单次屏气电影法MRI评价左心室功能的初步研究[J].中国医学影像技术,2005,21(1):65-67. 被引量:12
  • 2Schreiber W, Vorwerk CK, Langenbucher A, et al. A comparison of rebound tonometer(ICARE) with tonometer and GAT. Ophthalmologe 2007;104(4):299-304
  • 3Sahin A, Basmak H, Niyaz L, et al. Reproducibility and tolerability of the ICARE rebound tonometer in school children. J Glaucoma 2007;16(2):185-188
  • 4Toyohara K, Inoue K, et al. Comparison of ICARE rebound tonometer with GAT and non-contact tonometer. Atarashii (Journal of the Eye) 2007;24(3):355-359
  • 5López-Caballero C, Contreras I, Muos-Negrete FJ, et al. Rebound tonometry in a clinical setting.Comparison with applanation tonometry. Esp Oftalmol 2007;82(5):273-278
  • 6Queirós A, González-Méijome JM, Fernandes P, et al. A comparison of central and peripheral intraocular pressure using rebound tonometry. Ophthal Physiol Opt 2007;27(5):506-511
  • 7Davies LN, Bartlett H, Mallen EA, et al. Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 2006;84(2):206-209
  • 8傅伟才,张琳.开角型青光眼患者昼夜眼压监测及其意义[J].上海交通大学学报(医学版),2007,27(9):1163-1165. 被引量:3
  • 9Bland JM,Altman DG.Statistical methods for assessing agreement between two methods of clinical measurement.Lancet,1986,i:307-310.
  • 10Bland JM,Altman DG.Measuring agreement in method comparison studies.Statis tical Methodsin Medical Research,1999,8:135-160.

共引文献232

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部