期刊文献+

CT增强对比剂防渗漏导流报警系统研发的实验研究 被引量:1

An anti-leak stream diversion alarm system for contrast agent in CT enhancement scanning
下载PDF
导出
摘要 目的研发一种能辅助CT增强扫描高压注射器及时、安全和准确杜绝对比剂外渗的导流报警系统,以有效避免并发症的发生。方法利用新型的压力传感器模块瞬时感应皮肤张力的变化,通过自毁式导流阀,准确实施对比剂外渗后的警示、导流和自毁功能,将对比剂外渗的危害控制在渗漏初期。并采用动物实验检测该系统的及时性、准确性和安全性。结果通过动物实验,有效验证了报警系统在高压注射器对比剂外渗的初期可瞬时、准确报警,及时切换通道实现导流、自毁。结论该系统设备结构简单、技术可靠,可及时、有效防止高压注射器对比剂外渗对肢体的损害,具有较好的临床应用前景。 Objective To develop a stream diverting alarm device which can aid promptly, safely and accurately in preventing leakage of high pressure injectors of contrast media for CT enhancement scanning. Methods A new type of pressure sensor module was utilized to mo- nitor skin tension in real time, and prepared to initiate alarm, divert stream and self-destruct by a self-destructive valve, thus limiting the po- tential hazard of leakage in its early stage. Animal tests were conducted to verify its prompt, accurate and safe function. Results The device has been validated in animal tests, which it is capable of providing instant and accurate alarm, prompt channel switching for diversion and self- destruct. Conclusion The device is simple in structure,with validated technology to promptly and effectively prevent damage to limb from leakage of high pressure injectors,which was of great perspective in clinical setting.
出处 《局解手术学杂志》 2012年第6期587-589,共3页 Journal of Regional Anatomy and Operative Surgery
基金 重庆市科技攻关计划项目(CSTC2009AC5021)
关键词 计算机断层扫描 高压注射器 对比剂 导流报警系统 computed tomagraphy high pressure injector contrast agent stream diversion alarm system
  • 相关文献

参考文献10

  • 1陈玉娟,王士芹,赵梅玉.CT增强扫描对比剂渗漏原因分析及预防[J].护理实践与研究,2009,6(24):23-25. 被引量:12
  • 2Namasivayam S, Kalra MK,Torres WE, et al. Adverse reactions to intra- venous iodinated contrast media: an update[J]. Curr Probl Diagn Radi- ol,2006,35(4) :164 - 169.
  • 3Federle MP, Chang PJ, Confer S, et al. Frequency and effects of extrava- sation of nonionic contrast media during rapid bolus injection rates[ J].Radiology, 1998,206 (3) :637 - 640.
  • 4Kingston RJ,Young N,Sindhusake DP,et al. Study of patients with in- travenous contrast extravasation on CT studies, with radiology staff and ward staff canuulations[ J]. J Med Imaging Radiat Oncol,2012,56(2) : 163 - 167.
  • 5Wienbeck S, Fischbach R, Kloska SP,et al. Prospective study of access- site complication of automated contrast injection with peripheral venous access in MDCT[ J]. Am J Roentgenol,2010,195(4) :825 -829.
  • 6Bellin MF, Jakobsen JA,Tomassin I, et al. Contrast medium extravasation injury : guidelines for prevention and management [ J ]. Eur Radiol,2002, 12( 11 ) :2807 - 2812.
  • 7Juchem BC, Dallguol CM. Immediate adverse reactions to intravenous iodinated contrast media in computed tomography [ J]. Rev Lat Am En- fermagem,2007,15 ( 1 ) :78 - 83.
  • 8Cohan RH, Bullard MA, Ellis JH, et al. Local reactions after injection of iodinated contrast material: Detection, management and outcome [ J ]. Ac- ta Radiol, 1997,4( 11 ) :811 -818.
  • 9Powell CC, Li JM, Rodino L, et al. A new device to limit extravasation during contrast -enhanced CT [ J ]. Am J Roentgeno1,2000,174 ( 2 ) : 315 -318.
  • 10Amaral JG,Traubiei J,David GB,et al. Safety of power injector use in children as measured by incidence of extravasation [ J ]. Am J Roent- genol,2006,187(2) :580 -583. Amaral JG,Traubiei J,David GB,et al. Safety of power injector use in children as measured by incidence of extravasation [ J ]. Am J Roent- genol,2006,187(2) :580 -583.

二级参考文献8

共引文献11

同被引文献4

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部