摘要
论证了测定氟化物时当氟电极洗涤的最大空白值不同和电极电位漂移的情况下 。
The influence of the difference between electrode blank potential and the potential drift on the determination of fluoride by fluoride selective electrode is investigated. An one point calibration method is proposed to ensure the stability, reproducibility and accuracy in the sample analysis.
出处
《工业水处理》
CAS
CSCD
2000年第7期37-39,共3页
Industrial Water Treatment
关键词
氟电极
一点校正法
离子选择电极
水质监测
fluoride electrode
blank potential
potential drift
one point calibration